Repeated thermo-sterilisation further affects the reliability of positive end-expiratory pressure valves.

Julia Christine Hartung1, Georg Schmölzer, Gerd Schmalisch, Charles Christoph Roehr.   

Abstract

AIM: Positive end-expiratory pressure (PEEP) valves are used together with self-inflating bags (SIB) to provide a preset PEEP during manual ventilation. It has recently been shown that these valves deliver highly variable levels of PEEP. We hypothesised that material fatigue due to repeated thermo-sterilisation (TS) may contribute to varying reliability of PEEP valves.
METHODS: In a laboratory study 10 new PEEP valves were tested before and after 10, 20 and 30 cycles of routine TS (7 min at 134°C) by using a neonatal lung model (compliance 0.2 mL/kPa). Settings were positive inflation pressure = 20 and 40 cm H(2)O, PEEP = 5 and 10 cm H(2)O, respiratory rate = 40 and 60/min, flow = 8l/min. PEEP was recorded using a respiratory function monitor.
RESULTS: Before TS, a mean (standard deviation) PEEP of 4.0 (0.9) and 7.7 (1.0) cm H(2)O was delivered by the 10 valves when the PEEP was set to 5 and 10 cm H(2)O, respectively. One new valve only delivered 2.0 (0.0) and 5.0 (0.0) cm H(2)O when the PEEP was adjusted to 5 and 10 cm H(2)O, respectively. Four of the 10 investigated valves showed significant variations in PEEP (coefficient of variation >10%) throughout the autoclaving process. One valve completely lost its function after the 20th TS. Common defects were tears in the softer materials or displacement of the rubber seal. Six of the 10 valves continued to provide PEEP in spite of repeated TS.
CONCLUSION: The reliability of PEEP valves is affected by repeated TS. Multi-use PEEP valves should be tested for reliable PEEP provision following TS.
© 2013 The Authors. Journal of Paediatrics and Child Health © 2013 Paediatrics and Child Health Division (Royal Australasian College of Physicians).

Entities:  

Keywords:  PEEP; airway; neonate; resuscitation; valve.

Mesh:

Year:  2013        PMID: 23731063     DOI: 10.1111/jpc.12258

Source DB:  PubMed          Journal:  J Paediatr Child Health        ISSN: 1034-4810            Impact factor:   1.954


  2 in total

1.  Reliability of Single-Use PEEP-Valves Attached to Self-Inflating Bags during Manual Ventilation of Neonates--An In Vitro Study.

Authors:  Julia C Hartung; Silke Wilitzki; Marta Thio-Lluch; Arjan B te Pas; Gerd Schmalisch; Charles C Roehr
Journal:  PLoS One       Date:  2016-02-25       Impact factor: 3.240

Review 2.  [Newborn resuscitation and support of transition of infants at birth].

Authors:  John Madar; Charles C Roehr; Sean Ainsworth; Hege Ersda; Colin Morley; Mario Rüdiger; Christiane Skåre; Tomasz Szczapa; Arjan Te Pas; Daniele Trevisanuto; Berndt Urlesberger; Dominic Wilkinson; Jonathan P Wyllie
Journal:  Notf Rett Med       Date:  2021-06-02       Impact factor: 0.892

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.