Samuel A Spear1, James V Crawford.
Abstract
Objective. To visually compare the Nitinol "smart" stapes prosthesis to conventional manual crimping stapes pistons in temporal bone cadaver specimens. Main Outcome Measures. 10 otolaryngologists were given a photograph of the randomly ordered stapes pistons and asked to use the pictures to answer questions about each stapes piston. The answers to the survey were then recorded for analysis. Results. 8 of 9 Nitinol pistons were described as circular, and 3 of 9 manual crimped pistons were described as circular (P < .05). 6 of 9 Nitinol pistons were considered to be in contact with >66% of the incus and 3 of 9 to be in contact with 34-66% of the incus. 3 of 9 manually crimped pistons were considered to be in contact with >66% of the incus, 3 with 34-66% contact and 3 with less than 34% contact. Conclusions. The Nitinol "smart" stapes pistons were considered to provide a more circular and circumferential crimping and to have greater contact with the long process of the incus than conventional stapes pistons.Entities:
Year: 2011 PMID: 23724262 PMCID: PMC3658642 DOI: 10.5402/2011/932849
Source DB: PubMed Journal: ISRN Otolaryngol ISSN: 2090-5742
Figure 1Pictures of the 18 crimped stapes pistons after dissection out of the temporal bone specimens. Conventional manual crimped pistons: (a), (b), (f), (g), (h), (l), (m), (o), and (p). Nitinol “smart” pistons: (c), (d), (e), (i), (j), (k), (n), (q), and (r).
Survey results for the conventional manual crimped stapes prosthesis (%), n = 10 (survey respondents).
| Piston Letter | Considered | Considered | Considered circular | Considered oval | Considered to have <34% contact with the incus | Considered to have 33–46% contact with the incus | Considered to have >66% contact with the incus |
|---|---|---|---|---|---|---|---|
| (a) | 60% | 40% | 90% | 10% | 40% | 0% | 50% |
| (b) | 30% | 70% | 0% | 100% | 10% | 50% | 40% |
| (f) | 30% | 70% | 20% | 80% | 10% | 60% | 30% |
| (g) | 50% | 50% | 40% | 60% | 50% | 40% | 10% |
| (h) | 40% | 60% | 80% | 20% | 0% | 20% | 80% |
| (l) | 30% | 70% | 10% | 90% | 60% | 20% | 20% |
| (m) | 30% | 70% | 0% | 100% | 20% | 50% | 30% |
| (o) | 50% | 50% | 90% | 10% | 30% | 20% | 50% |
| (p) | 30% | 70% | 20% | 80% | 80% | 20% | 0% |
Survey results for the Nitinol SMART crimped stapes prosthesis (%), n = 10 (survey respondents).
| Piston Letter | Considered | Considered | Considered circular | Considered oval | Considered to have <34% contact with the incus | Considered to have 33–46% contact with the incus | Considered to have >66% contact with the incus |
|---|---|---|---|---|---|---|---|
| (c) | 40% | 60% | 40% | 60% | 20% | 50% | 30% |
| (d) | 60% | 40% | 100% | 0% | 10% | 20% | 70% |
| (e) | 50% | 50% | 80% | 20% | 10% | 40% | 50% |
| (i) | 60% | 40% | 90% | 10% | 10% | 20% | 70% |
| (j) | 50% | 50% | 80% | 20% | 20% | 0% | 80% |
| (k) | 50% | 50% | 90% | 10% | 10% | 10% | 80% |
| (n) | 50% | 50% | 70% | 30% | 40% | 60% | 0% |
| (q) | 50% | 50% | 90% | 10% | 10% | 30% | 60% |
| (r) | 50% | 50% | 90% | 10% | 10% | 50% | 40% |
Survey results for the Nitinol SMART crimped stapes prosthesis (%), n = 9 (total of each type of piston).
| Piston Type | # Pistons considered circular by ≥50% of respondents, | # Pistons considered oval by ≥50% of respondents, |
| # of pistons considered to have <34% contact with the incus by ≥50% of respondents | # of pistons considered to have 34–66% contact with the incus by ≥50% of respondents | # of pistons considered to have <66% contact with the incus by ≥50% of respondents |
|
|---|---|---|---|---|---|---|---|
| Conventional | 3 (33%) | 6 (66%) | <0.05 | 3 (33%) | 3 (33%) | 3 (33%) | 0.35 |
| Nitinol | 8 (88.9%) | 1 (11.1%) | 0 | 3 (33%) | 6 (66%) |