Literature DB >> 23720556

Round Afghanistan with a fridge.

Gareth Wild, D Anderson, P Lund.   

Abstract

INTRODUCTION: This paper covers the contemporary deployment of blood products in the pre-hospital environment during extended field operations in Afghanistan. The equipment used was standard-issue to the British Armed Forces but used in a novel manner. The aim of this paper was to establish the reliability of the RCB42P blood bank and the concept of blood storage in the field during protracted vehicle-borne patrols.
METHOD: TempIT tag data was collected for five patrols and analysed. Ambient temperatures varied immensely from minus 5°C to plus 50°C. Mitigation measures were introduced to reduce the exposure of the blood bank to radiated and ambient heat.
RESULTS: The blood bank is affected by radiated heat from the engine compartment on long journeys coupled with high ambient temperatures. However, this can be minimised by simple mitigation measures using insulation and passive cooling.
CONCLUSIONS: The RCB42P blood bank is a robust unit capable of enduring a considerable amount of physical abuse and extremes of temperature. It is suitable for the storage of blood products on protracted vehicle-borne patrols in high ambient temperatures if the mitigation measures outlined are employed.

Entities:  

Year:  2013        PMID: 23720556     DOI: 10.1136/jramc-2013-000013

Source DB:  PubMed          Journal:  J R Army Med Corps        ISSN: 0035-8665            Impact factor:   1.285


  1 in total

Review 1.  Prehospital Blood Product Resuscitation for Trauma: A Systematic Review.

Authors:  Iain M Smith; Robert H James; Janine Dretzke; Mark J Midwinter
Journal:  Shock       Date:  2016-07       Impact factor: 3.454

  1 in total