Experimental analysis of general ion recombination in a liquid-filled ionization chamber in high-energy photon beams.

Eunah Chung1, Stephen Davis, Jan Seuntjens.   

Abstract

PURPOSE: To study experimentally the general ion recombination effect in a liquid-filled ionization chamber (LIC) in high-energy photon beams.
METHODS: The general ion recombination effect on the response of a micro liquid ion chamber (microLion) was investigated with a 6 MV photon beam in normal and SRS modes produced from a Varian(®) Novalis Tx(TM) linear accelerator. Dose rates of the linear accelerator were set to 100, 400, and 1000 MU∕min, which correspond to pulse repetition frequencies of 60, 240, and 600 Hz, respectively. Polarization voltages applied to the microLion were +800 and +400 V. The relative collection efficiency of the microLion response as a function of dose per pulse was experimentally measured with changing polarization voltage and pulse repetition frequencies and was compared with the theoretically calculated value.
RESULTS: For the 60 Hz pulse repetition frequency, the experimental relative collection efficiency was not different from the theoretical one for a pulsed beam more than 0.3% for both polarization voltages. For a pulsed radiation beam with a higher pulse repetition frequency, the experimental relative collection efficiency converged to the theoretically calculated efficiency for continuous beams. This result indicates that the response of the microLion tends toward the response to a continuous beam with increasing pulse repetition frequency of a pulsed beam because of low ion mobility in the liquid.
CONCLUSIONS: This work suggests an empirical method to correct for differences in general ion recombination of a LIC between different radiation fields. More work is needed to quantitatively explain the LIC general ion recombination behavior in pulsed beams generated from linear accelerators.

Mesh:

Substances:

Year:  2013        PMID: 23718602     DOI: 10.1118/1.4805109

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  General ion recombination effect in a liquid ionization chamber in high-dose-rate pulsed photon and electron beams.

Authors:  Hideyuki Takei; Tomonori Isobe; Nozomi Kitamura; Yutaro Mori; Tetsuya Tomita; Daisuke Kobayashi; Satoshi Kamizawa; Tomoharu Sato; Hideyuki Sakurai; Takeji Sakae
Journal:  J Radiat Res       Date:  2018-05-01       Impact factor: 2.724