Changes realized from extended bit-depth and metal artifact reduction in CT.

C Glide-Hurst1, D Chen, H Zhong, I J Chetty.   

Abstract

PURPOSE: High-Z material in computed tomography (CT) yields metal artifacts that degrade image quality and may cause substantial errors in dose calculation. This study couples a metal artifact reduction (MAR) algorithm with enhanced 16-bit depth (vs standard 12-bit) to quantify potential gains in image quality and dosimetry.
METHODS: Extended CT to electron density (CT-ED) curves were derived from a tissue characterization phantom with titanium and stainless steel inserts scanned at 90-140 kVp for 12- and 16-bit reconstructions. MAR was applied to sinogram data (Brilliance BigBore CT scanner, Philips Healthcare, v.3.5). Monte Carlo simulation (MC-SIM) was performed on a simulated double hip prostheses case (Cerrobend rods embedded in a pelvic phantom) using BEAMnrc∕Dosxyz (400,000,0000 histories, 6X, 10 × 10 cm(2) beam traversing Cerrobend rod). A phantom study was also conducted using a stainless steel rod embedded in solid water, and dosimetric verification was performed with Gafchromic film analysis (absolute difference and gamma analysis, 2% dose and 2 mm distance to agreement) for plans calculated with Anisotropic Analytic Algorithm (AAA, Eclipse v11.0) to elucidate changes between 12- and 16-bit data. Three patients (bony metastases to the femur and humerus, and a prostate cancer case) with metal implants were reconstructed using both bit depths, with dose calculated using AAA and derived CT-ED curves. Planar dose distributions were assessed via matrix analyses and using gamma criteria of 2%∕2 mm.
RESULTS: For 12-bit images, CT numbers for titanium and stainless steel saturated at 3071 Hounsfield units (HU), whereas for 16-bit depth, mean CT numbers were much larger (e.g., titanium and stainless steel yielded HU of 8066.5 ± 56.6 and 13,588.5 ± 198.8 for 16-bit uncorrected scans at 120 kVp, respectively). MC-SIM was well-matched between 12- and 16-bit images except downstream of the Cerrobend rod, where 16-bit dose was ∼6.4% greater than 12-bit. Absolute film dosimetry in a region downstream of a stainless steel rod revealed that 16-bit calculated dose, with and without MAR, agreed more closely with film results (1%-2% less than film) as compared to 12-bit reconstructions (5.6%-6.5% less than film measurements). Gamma analysis revealed that 16-bit dose calculations were better matched to film results than 12-bit (∼10% higher pass rates for 16-bit). Similar results were observed in two patient cases; the largest discrepancy was observed for a femur case where 12-bit doses, both with and without MAR correction, were 6-7 Gy lower (∼17%-20% of the prescription dose) as compared to 16-bit dose calculations. However, when beams are not directly traversing metal, such as a prostate cancer case with bilateral hip prostheses; the impact of 16-bit reconstruction was diminished.
CONCLUSIONS: These results suggest that it may be desirable to implement 16-bit MAR-corrected images for treatment planning purposes, which can provide a more accurate dosimetric approach coupled with improved visualization by suppression of CT artifacts.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23718590     DOI: 10.1118/1.4805102

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Development and first validation of a simplified CT-based classification system of soft tissue changes in large-head metal-on-metal total hip replacement: intra- and interrater reliability and association with revision rates in a uniform cohort of 664 arthroplasties.

Authors:  Martijn F Boomsma; Mireille A Edens; Christiaan P Van Lingen; Niek Warringa; Harmen B Ettema; Cees C P M Verheyen; Mario Maas
Journal:  Skeletal Radiol       Date:  2015-05-06       Impact factor: 2.199

2.  An evaluation of three commercially available metal artifact reduction methods for CT imaging.

Authors:  Jessie Y Huang; James R Kerns; Jessica L Nute; Xinming Liu; Peter A Balter; Francesco C Stingo; David S Followill; Dragan Mirkovic; Rebecca M Howell; Stephen F Kry
Journal:  Phys Med Biol       Date:  2015-01-14       Impact factor: 3.609

3.  Iterative metal artifact reduction improves dose calculation accuracy : Phantom study with dental implants.

Authors:  Manuel Maerz; Pia Mittermair; Andreas Krauss; Oliver Koelbl; Barbara Dobler
Journal:  Strahlenther Onkol       Date:  2016-03-11       Impact factor: 3.621

4.  Application of Metal Implant 16-Bit Imaging: New Technique in Radiotherapy.

Authors:  Ni Xin-Ye; Gao Liugang; Fang Mingming; Lin Tao
Journal:  Technol Cancer Res Treat       Date:  2016-06-23

5.  Approaches to reducing photon dose calculation errors near metal implants.

Authors:  Jessie Y Huang; David S Followill; Rebecca M Howell; Xinming Liu; Dragan Mirkovic; Francesco C Stingo; Stephen F Kry
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

6.  Cone beam CT for determining breast cancer margin: an initial experience and its comparison with mammography and specimen radiograph.

Authors:  Juan Yao; Chris Shaw; C J Lai; John Rong; Jian Wang; Wenya Liu
Journal:  Int J Clin Exp Med       Date:  2015-09-15

7.  The effects of the orthopedic metal artifact reduction (O-MAR) algorithm on contouring and dosimetry of head and neck radiotherapy patients.

Authors:  Jussi Sillanpaa; Michael Lovelock; Boris Mueller
Journal:  Med Dosim       Date:  2019-07-30       Impact factor: 1.482

Review 8.  Computed tomography imaging parameters for inhomogeneity correction in radiation treatment planning.

Authors:  Indra J Das; Chee-Wai Cheng; Minsong Cao; Peter A S Johnstone
Journal:  J Med Phys       Date:  2016 Jan-Mar

9.  Clinically applicable artificial intelligence system for dental diagnosis with CBCT.

Authors:  Matvey Ezhov; Maxim Gusarev; Maria Golitsyna; Julian M Yates; Evgeny Kushnerev; Dania Tamimi; Secil Aksoy; Eugene Shumilov; Alex Sanders; Kaan Orhan
Journal:  Sci Rep       Date:  2021-07-22       Impact factor: 4.379

10.  Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study.

Authors:  Ching-Ching Yang; Fong-Lin Chen; Yeh-Chi Lo
Journal:  PLoS One       Date:  2016-06-09       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.