Evaluation of the redesigned conformable GORE TAG thoracic endoprosthesis for traumatic aortic transection.

Mark A Farber1, Joseph S Giglia, Benjamin W Starnes, Scott L Stevens, Jeremiah Holleman, Rabih Chaer, Jon S Matsumura.   

Abstract

OBJECTIVE: To evaluate the safety and effectiveness of the conformable GORE TAG thoracic endoprosthesis (CTAG) device (W. L. Gore and Associates, Flagstaff, Ariz) for the endovascular repair of traumatic aortic transections.
METHODS: A prospective, nonrandomized, multicenter trial was conducted at 21 sites. Primary safety end points included 30-day all-cause mortality. The effectiveness end point was freedom from a major device event requiring reintervention through 1-month follow-up.
RESULTS: Fifty-one subjects were enrolled between December 2009 and January 2011 with polytraumatic injuries and a mean Injury Severity Score of 32 ± 14. The proximal mean intimal aortic diameter measured 24 mm, while the mean distal intimal diameter was 22 mm. A total of 57 CTAG devices were implanted (mean, 1.1/subject; range, 1-2) with a mean patient age of 44 years (range, 21-87) and a male-to-female ratio of 2:1. Technical success was 100% with an operative mortality of 0%. Femoral access was utilized in 96% of patients. The mean procedure time and blood loss was 105 minutes and 148 mL, respectively. All subjects required admission to an intensive care unit with a mean hospital stay of 14.6 days. Adjuvant techniques (ie, lumbar drains and induced hypertension) to prevent paraplegia were used in only 7.8% of patients. No patient developed paraplegia despite 63% having complete or partial left subclavian artery coverage and only 9% of those receiving left subclavian artery revascularization. In addition, there were no device compressions or major device events reported. Overall mortality at 30 days was 7.8%, and all were adjudicated by the clinical events committee as not being device or procedure related. Serious adverse events occurred in 39.2% of patients through 30 days. To date, there have been no conversions to open repair. Two site-reported endoleaks were detected during the mean follow-up of 4.2 months, which did not require reintervention.
CONCLUSIONS: The CTAG device was demonstrated to be a safe and effective treatment for traumatic aortic transection based on 30-day outcomes. There were no device-related serious adverse events.
Copyright © 2013 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23711695     DOI: 10.1016/j.jvs.2013.02.015

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  4 in total

1.  Aortic morphometry at endograft position as assessed by 3D image analysis affects risk of type I endoleak formation after TEVAR.

Authors:  Drosos Kotelis; Carolin Brenke; Stefan Wörz; Fabian Rengier; Karl Rohr; Hans-Ulrich Kauczor; Dittmar Böckler; Hendrik von Tengg-Kobligk
Journal:  Langenbecks Arch Surg       Date:  2015-02-22       Impact factor: 3.445

2.  Contemporary evaluation of mortality and stroke risk after thoracic endovascular aortic repair.

Authors:  Frances Y Hu; Zachary B Fang; Bradley G Leshnower; Yazan Duwayri; William D Jordan; Theresa W Gillespie; Ravi K Veeraswamy
Journal:  J Vasc Surg       Date:  2017-05-11       Impact factor: 4.268

3.  A 10-Year Single-Center Experience With the GORE TAG Conformable Thoracic Stent Graft in the Treatment of Thoracic Aortic Disease.

Authors:  Denis Skrypnik; Moritz S Bischoff; Katrin Meisenbacher; Dorothea B Kronsteiner; Dittmar Böckler
Journal:  J Endovasc Ther       Date:  2021-10-11       Impact factor: 3.089

4.  Device Conformability and Morphological Assessment After TEVAR for Aortic Type B Dissection: A Single-Centre Experience with a Conformable Thoracic Stent-Graft Design.

Authors:  Moritz S Bischoff; Matthias Müller-Eschner; Katrin Meisenbacher; Andreas S Peters; Dittmar Böckler
Journal:  Med Sci Monit Basic Res       Date:  2015-12-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.