Outcomes of torsional microcoaxial phacoemulsification performed by 12-degree and 22-degree bent tips.

Firat Helvacioglu1, Celal Yeter, Zeki Tunc, Sadik Sencan.   

Abstract

PURPOSE: To compare the safety and efficacy of Ozil Intelligent Phaco torsional microcoaxial phacoemulsification surgeries performed with 12-degree and 22-degree bent tips using the Infiniti Vision System.
SETTING: Maltepe University School of Medicine Department of Ophthalmology, Istanbul, Turkey.
DESIGN: Comparative case series.
METHODS: Eyes were assigned to 2.2 mm microcoaxial phacoemulsification using the torsional mode with a 22-degree bent tip (Group 1) or a 12-degree bent tip (Group 2). The primary outcome measures were ultrasound time (UST), cumulative dissipated energy (CDE), longitudinal and torsional ultrasound (US) amplitudes, mean surgical time, mean volume of balanced salt solution used, and surgical complications.
RESULTS: Both groups included 45 eyes. The mean UST, CDE, longitudinal US amplitude, and torsional US amplitude were 65 seconds ± 27.23 (SD), 11.53 ± 6.99, 0.22 ± 0.26, and 42.86 ± 15.64, respectively, in Group 1 and 84 ± 45.04 seconds, 16.68 ± 10.66, 0.48 ± 0.68, and 46.27 ± 14.74, respectively, in Group 2. The mean UST, CDE, and longitudinal amplitudes were significantly lower in Group 1 (P=.003, P=.008, and P=.022, respectively). The mean volume of balanced salt solution was 73.33 ± 28.58 cc in Group 1 and 82.08 ± 26.21 cc in Group 2 (P=.134).
CONCLUSION: Torsional phacoemulsification performed with 22-degree bent tips provided more effective lens removal than 12-degree bent tips, with a lower UST and CDE.
Copyright © 2013 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23707410     DOI: 10.1016/j.jcrs.2013.03.002

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Efficacy of Balanced Torsional Phacoemulsification Tip for Cataract Surgery.

Authors:  Huri Sabur; Sayime Aydin Eroglu; Salih Sertac Azarsiz
Journal:  J Curr Ophthalmol       Date:  2022-04-16

2.  Results of endocapsular phacofracture debulking of hard cataracts.

Authors:  James A Davison
Journal:  Clin Ophthalmol       Date:  2015-07-10

3.  Effect of IOP based infusion system with and without balanced phaco tip on cumulative dissipated energy and estimated fluid usage in comparison to gravity fed infusion in torsional phacoemulsification.

Authors:  Praveen K Malik; Taru Dewan; Arun Kr Patidar; Ekta Sain
Journal:  Eye Vis (Lond)       Date:  2017-09-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.