Liver iron quantification by 3 tesla MRI: calibration on a rabbit model.

Peng Peng1, Zhongkui Huang, Liling Long, Fanyu Zhao, Chunyan Li, Wenmei Li, Taigang He.   

Abstract

PURPOSE: To determine the feasibility of liver iron quantification by 3 Tesla (T) MRI using a novel iron overload rabbit model.
MATERIALS AND METHODS: Forty-two rabbits underwent iron dextran loading from 1 to 15 weeks. MRI signal intensity ratio (SIR) was measured using a gradient-echo sequence, and R2(1/T2) measured using an eight-echo spin-echo sequence at 3T. Ex vivo hepatic pathology was obtained for all rabbits studied. Postmortem assessments of liver iron concentration (LIC) were conducted in an atomic absorption spectrophotometer. MRI measures were fitted against LIC using linear regression for 30 of the iron-loaded rabbits. The remaining 12 iron-loaded rabbits were used to test the prediction accuracy of the derived models.
RESULTS: LIC was linearly correlated to both liver-to-muscle SIR (r = -0.845) and R2 (r = 0.965) in a range achieved in this study (LIC < 10 mg/g dry tissue) at 3T. By regression, the linear equations were determined as: Y1 = 10.581-5.924X1 (Y1 : LIC, X1 :SIR); Y2 = -1.273+0.103X2 (Y2 :LIC, X2 :R2). In the 12 test rabbits, the predicted LICs using the derived equations agreed well with the results obtained using the spectrophotometer.
CONCLUSION: Both SIR and R2 are highly correlated with LIC in a novel rabbit model. MRI quantification of liver iron overload is feasible at 3T.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  3T; MRI; iron overload; liver; quantification; rabbit model

Mesh:

Substances:

Year:  2013        PMID: 23704041     DOI: 10.1002/jmri.24074

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  Assessment of liver iron overload by 3 T MRI.

Authors:  A Paisant; A Boulic; E Bardou-Jacquet; E Bannier; G d'Assignies; F Lainé; B Turlin; Y Gandon
Journal:  Abdom Radiol (NY)       Date:  2017-06

2.  Relaxivity-iron calibration in hepatic iron overload: Reproducibility and extension of a Monte Carlo model.

Authors:  Changqing Wang; Scott B Reeder; Diego Hernando
Journal:  NMR Biomed       Date:  2021-08-30       Impact factor: 4.044

3.  The application value of ultra-short echo time MRI in the quantification of liver iron overload in a rat model.

Authors:  Qiaoling Wu; Xiuwei Fu; Zhizheng Zhuo; Mingfeng Zhao; Hongyan Ni
Journal:  Quant Imaging Med Surg       Date:  2019-02

4.  Design and construction of an optimized transmit/receive hybrid birdcage resonator to improve full body images of medium-sized animals in 7T scanner.

Authors:  Nibardo Lopez Rios; Philippe Pouliot; Konstantinos Papoutsis; Alexandru Foias; Nikola Stikov; Frédéric Lesage; Mathieu Dehaes; Julien Cohen-Adad
Journal:  PLoS One       Date:  2018-02-01       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.