Inertance estimation in a lumped-parameter hydraulic simulator of human circulation.

Ettore Lanzarone1, Fabrizio Ruggeri.   

Abstract

Pulsatile mock loop systems are largely used to investigate the cardiovascular system in vitro. They consist of a pump, which replicates the heart, coupled with a lumped-parameter hydraulic afterload, which simulates vasculature. An accurate dimensioning of components is required for a reliable mimicking of the physiopathological behavior of the system. However, it is not possible to create a component for the afterload inertance, and inertance contributions are present in the entire circuit. Hence, in the literature, inertance is neglected or qualitatively evaluated. In this paper, we propose two quantitative methods (Maximum-likelihood estimation (MLE) and Bayesian estimation) for estimating afterload inertance based on observed pressure and flow waveforms. These methods are also applied to a real mock loop system. Results show that the system has an inertance comparable with the literature reference value of the entire systemic circulation, and that the expected variations over inlet average flow and pulse frequency are in general confirmed. Comparing the methods, the Bayesian approach results in higher and more stable estimations than the classical MLE.

Entities:  

Mesh:

Year:  2013        PMID: 23699724     DOI: 10.1115/1.4024138

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  1 in total

1.  A Zero-Dimensional Model and Protocol for Simulating Patient-Specific Pulmonary Hemodynamics From Limited Clinical Data.

Authors:  Vitaly O Kheyfets; Jamie Dunning; Uyen Truong; Dunbar Ivy; Kendall Hunter; Robin Shandas
Journal:  J Biomech Eng       Date:  2016-12-01       Impact factor: 2.097

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.