Mechanical properties of bioresorbable self-reinforced posterior cervical rods.

Katherine Savage1, Zeeshan M Sardar, Timo Pohjonen, Gursukhman S Sidhu, Benjamin D Eachus, Alexander Vaccaro.   

Abstract

STUDY
DESIGN: A biomechanical study.
OBJECTIVE: To test the mechanical and physical properties of self-reinforced copolymer bioresorbable posterior cervical rods and compare their mechanical properties to commonly used Irene titanium alloy rods. SUMMARY OF BACKGROUND DATA: Bioresorbable instrumentation is becoming increasingly common in surgical spine procedures. Compared with metallic implants, bioresorbable implants are gradually reabsorbed as the bone heals, transferring the load from the instrumentation to bone, eliminating the need for hardware removal. In addition, bioresorbable implants produce less stress shielding due to a more physiological modulus of elasticity.
METHODS: Three types of rods were used: (1) 5.5 mm copolymer rods and (2) 3.5 mm and (3) 5.5 mm titanium alloy rods. Four tests were used on each rod: (1) 3-point bending test, (2) 4-point bending test, (3) shear test, and (4) differential scanning calorimeter test. The outcomes were recorded: Young modulus (E), stiffness, maximum load, deflection at maximum load, load at 1.0% strain of the rod's outer surface, and maximum bending stress.
RESULTS: The Young modulus (E) for the copolymer rods (mean range, 6.4-6.8 GPa) was significantly lower than the 3.5 mm titanium rods (106 GPa) and the 5.5 mm titanium rods (95 GPa). The stiffness of the copolymer rods (mean range, 16.6-21.4 N/mm) was also significantly lower than the 3.5 mm titanium alloy rods (43.6 N/mm) and the 5.5 mm titanium alloy rods (239.6 N/mm). The mean maximum shear load of the copolymer rods was 2735 N and they had significantly lower mean maximum loads than the titanium rods.
CONCLUSIONS: Copolymer rods have adequate shear resistance, but less load resistance and stiffness compared with titanium rods. Their stiffness is closer to that of bone, causing less stress shielding and better gradual dynamic loading. Their use in semirigid posterior stabilization of the cervical spine may be considered.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 23698109     DOI: 10.1097/BSD.0b013e318299c6d8

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  2 in total

1.  Assessment of the suitability of biodegradable rods for use in posterior lumbar fusion: An in-vitro biomechanical evaluation and finite element analysis.

Authors:  Fon-Yih Tsuang; Yueh-Ying Hsieh; Yi-Jie Kuo; Chia-Hsien Chen; Feng-Huei Lin; Chen-Sheng Chen; Chang-Jung Chiang
Journal:  PLoS One       Date:  2017-11-16       Impact factor: 3.240

2.  Biomechanical analysis of single-level interbody fusion with different internal fixation rod materials: a finite element analysis.

Authors:  Yueh-Ying Hsieh; Fon-Yih Tsuang; Yi-Jie Kuo; Chia-Hsien Chen; Chang-Jung Chiang; Chun-Li Lin
Journal:  BMC Musculoskelet Disord       Date:  2020-02-14       Impact factor: 2.362

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.