| Literature DB >> 23690775 |
Paola Figueroa-Barojas1, Mihir R Bakhru, Nagy A Habib, Kristi Ellen, Jennifer Millman, Armeen Jamal-Kabani, Monica Gaidhane, Michel Kahaleh.
Abstract
UNLABELLED: Objectives. Radiofrequency ablation (RFA) has replaced photodynamic therapy for premalignant and malignant lesions of the esophagus. However, there is limited experience in the bile duct. The objective of this pilot study was to assess the safety and efficacy of RFA in malignant biliary strictures.Entities:
Year: 2013 PMID: 23690775 PMCID: PMC3649248 DOI: 10.1155/2013/910897
Source DB: PubMed Journal: J Oncol ISSN: 1687-8450 Impact factor: 4.375
Figure 1Fluoroscopic images of bile duct cancer at the confluence with a Bismuth III lesion.
Figure 2EndoHPB Probe for radio frequency ablation.
Figure 3Application of radiofrequency at the level of the left hepatic duct.
Figure 4Application of radiofrequency ablation at the level of the right hepatic duct.
Figure 5Placement of biliary stent posts radiofrequency application.
Patient demographics.
| Number of patients | 20 |
| Age y, median (range) | 65.3 (45–86) |
| Sex (male/female) | 15/5 |
| Disease: | |
| Cholangiocarcinoma | 11 |
| Pancreatic cancer | 7 |
| IPMN | 1 |
| Gastric cancer | 1 |
Treatment results.
| Number of strictures treated | 25 |
| Mean stricture length (mm) | 15.2 (SD = 8.7, range = 3.5–33) |
| Mean stricture diameter before RFA (mm) | 1.7 (SD = 0.9, range = 0.5–3.4) |
| Mean stricture diameter after RFA (mm) | 5.2 (SD = 2, range = 2.6–9) |
| Type of stent placed | |
| Uncovered SEMS* | 1 |
| Partially or fully covered | 13 |
| Plastic stents | 6 |
| Complications | |
| Pain | 5 |
| Mild after ERCP | 1 |
| Cholecystitis | 1 |
*SEMS: self-expanding metal stent.