OBJECTIVE: Complex endovascular procedures such as fenestrated endovascular aneurysm repair (FEVAR) are associated with higher radiation doses compared with other fluoroscopically guided interventions (FGIs). The purpose of this study was to determine whether surgeon education on radiation dose control can lead to lower reference air kerma (RAK) and peak skin dose (PSD) levels in high-dose procedures.
METHODS: Radiation dose and operating factors were recorded for FGI performed in a hybrid room over a 16-month period. Cases exceeding 6 Gy RAK were investigated according to institutional policy. Information obtained from these investigations led to surgeon education focused on reducing patient dose. Points addressed included increasing table height, utilizing collimation and angulation, decreasing magnification modes, and maintaining minimal patient-to-detector distance. Procedural RAK doses and operating factors were compared 8 months pre- (group A) and 8 months post- (group B) educational intervention using analysis of variance with Tukey pairwise comparisons and t-tests. PSD distributions were calculated using custom software employing input data from fluoroscopic machine logs.
RESULTS: Of 447 procedures performed, 300 FGIs had sufficient data to be included in the analysis (54% lower extremity, 11% thoracic endovascular aneurysm repair, 10% cerebral, 8% FEVAR, 7% endovascular aneurysm repair, 5% visceral, and 5% embolization). Twenty-one cases were investigated for exceeding 6 Gy RAK. FEVAR comprised 70% of the investigated cases and had a significantly higher median RAK dose compared with all other FGIs (P < .0001). There was no difference in body mass index between groups A and B; however, increasing body mass index was an indicator for increased RAK. PSD calculations were performed for the 122 procedures that focused on the thorax and abdomen (group A, 80 patients; group B, 42 patients). Surgeon education most strongly affected table height, with an average table height elevation of 10 cm per case after education (P < .0001). The dose index (PSD/RAK ratio) was used to track changes in operating practices, and it decreased from 1.14 to 0.79 after education (P < .0001). These changes resulted in an estimated 16% reduction in PSD. There was a trend toward a decrease in patient to detector distance, and the use of collimation increased from 25% to 40% (P < .001) for all cases; however, these did not result in a decrease in PSD. The number of cases that exceeded 6 Gy RAK did not change after education; however, the proportion of non-FEVAR cases that exceeded 6 Gy decreased from 40% to 20%.
CONCLUSIONS: Surgeon education on the appropriate use of technical factors during FGIs improved operating practice, reduced patient radiation dose, and decreased the number of non-FEVAR cases that exceeded 6 Gy. It is essential that vascular surgeons be educated in best operating practices to lower PSD; nonetheless, FEVAR remains a high-dose procedure.
Copyright © 2013 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23688628     DOI: 10.1016/j.jvs.2013.04.004

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  10 in total

1.  Assessment of Radiation Safety Knowledge Among Urology Residents in the United States.

Authors:  Andrew M Harris; John Loomis; Marilyn Hopkins; Jason Bylund
Journal:  J Endourol       Date:  2019-05-17       Impact factor: 2.942

Review 2.  Physician and Patient Radiation Exposure During Endovascular Procedures.

Authors:  Andrew M Goldsweig; J Dawn Abbott; Herbert D Aronow
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-02

3.  Radiation dose during endovascular aneurysm repair (EVAR): upgrade of an angiographic system from standard to Eco mode.

Authors:  Maria Antonella Ruffino; Marco Fronda; Andrea Discalzi; Paola Isoardi; Laura Bergamasco; Roberto Ropolo; Dorico Righi; Paolo Fonio
Journal:  Radiol Med       Date:  2018-07-23       Impact factor: 3.469

4.  Three-dimensional fusion computed tomography decreases radiation exposure, procedure time, and contrast use during fenestrated endovascular aortic repair.

Authors:  Michael M McNally; Salvatore T Scali; Robert J Feezor; Daniel Neal; Thomas S Huber; Adam W Beck
Journal:  J Vasc Surg       Date:  2014-08-28       Impact factor: 4.268

5.  Transesophageal echocardiography as the sole guidance for occlusion of patent ductus arteriosus using a canine ductal occluder in dogs.

Authors:  F Porciello; D Caivano; M E Giorgi; P Knafelz; M Rishniw; N S Moise; A Bufalari; A Fruganti; F Birettoni
Journal:  J Vet Intern Med       Date:  2014-07-16       Impact factor: 3.333

6.  Fluoroscopy-induced chronic radiation dermatitis (FICRD) after endovascular abdominal aortic aneurysm endoleak repair.

Authors:  Alexis B Lyons; Valerie M Harvey; Julia Gusev
Journal:  JAAD Case Rep       Date:  2015-11-23

7.  Infectious complications of endoarterial interventional radiology: protocol for an observational study of a longitudinal national cohort of patients assessed in the French hospital discharge database (MOEVA study).

Authors:  Kaoutar Aid Mellouk; Abdelmajid Soulaymani; Fei Gao; Pascal Astagneau; Benoit Misset
Journal:  BMJ Open       Date:  2019-06-03       Impact factor: 2.692

Review 8.  Applications of Electrospun Drug-Eluting Nanofibers in Wound Healing: Current and Future Perspectives.

Authors:  Nakamwi Akombaetwa; Alick Bwanga; Pedzisai Anotida Makoni; Bwalya A Witika
Journal:  Polymers (Basel)       Date:  2022-07-20       Impact factor: 4.967

9.  The endovascular performance spectrum of vascular surgery departments in Germany: Results of an online survey among senior department physicians.

Authors:  T Schmitz-Rixen; G Torsello; M Steinbauer; R T Grundmann
Journal:  Gefasschirurgie       Date:  2016-08-03

10.  The hidden risk of ionizing radiation in the operating room: a survey among 258 orthopaedic surgeons in Brazil.

Authors:  Robinson Esteves Pires; Igor Guedes Nogueira Reis; Ângelo Ribeiro Vaz de Faria; Vincenzo Giordano; Pedro José Labronici; William Dias Belangero
Journal:  Patient Saf Surg       Date:  2020-04-22
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.