Non-invasive blood pressure and cardiac index measurements using the Finapres Portapres in an emergency department triage setting.

Yuri van der Does1, Lex M van Loon, Jelmer Alsma, Annebel Govers, Benno Lansdorp, Pleunie P M Rood, Stephanie C E Schuit.   

Abstract

UNLABELLED: Emergency department (ED) patients are triaged to determine the urgency of care. The Finapres Portapres (FP) measures blood pressure (BP) and cardiac output (CO) non-invasively, and may be of added value in early detection of patients at risk for hemodynamic compromise.
OBJECTIVES: Compare non-invasive BP measurements using FP and standard automated sphygmomanometry. Compare FP cardiac index (CI), CO corrected for body surface area, of normotensive patients, to chart-based physician estimate of shock, to discover if there is additional value in CI measurements in triage.
METHODS: ED Patients requiring BP measurement in triage were included. Systolic (SBP) and diastolic (DBP) BP were measured using both devices during a two minutes measurement. Two physicians independently judged probability of shock, defined as estimated CI ≤2.5 L min(-1) m(-2), based on chart review, three weeks after ED visit.
RESULTS: Of a total of 112 patients 97 patients were included. Pearson's correlation coefficient was 0.50 for SBP, 0.53 for DBP, with a Blant-Altman mean bias of 11.3 (upper limit 65.3, lower limit -42.8) and 7.7 (39.2, -23.7) for SBP and DBP respectively. In normotensive patients, the group with low FP CI measurements had significantly more cases with physician-estimated shock, compared to the normal to high measurements (P = .036).
CONCLUSIONS: When used as a triage device in the emergency department setting, non-invasive BP measurements using FP do not correlate well with automated sphygmomanometry. However, this study does indicate that use of the FP device in triage may aid physicians to recognize patients in early phases of shock.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23688570     DOI: 10.1016/j.ajem.2013.03.004

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


  2 in total

1.  Effects of tranexamic acid on coagulation indexes of patients undergoing heart valve replacement surgery under cardiopulmonary bypass.

Authors:  Fei Liu; Dong Xu; Kefeng Zhang; Jian Zhang
Journal:  Int J Immunopathol Pharmacol       Date:  2016-09-30       Impact factor: 3.219

2.  Limitations of stroke volume estimation by non-invasive blood pressure monitoring in hypergravity.

Authors:  Olivier Manen; Caroline Dussault; Fabien Sauvet; Stéphanie Montmerle-Borgdorff
Journal:  PLoS One       Date:  2015-03-23       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.