3-D residual eddy current field characterisation: applied to diffusion weighted magnetic resonance imaging.

Kieran O'Brien1, Alessandro Daducci, Nils Kickler, Francois Lazeyras, Rolf Gruetter, Thorsten Feiweier, Gunnar Krueger.   

Abstract

Clinical use of the Stejskal-Tanner diffusion weighted images is hampered by the geometric distortions that result from the large residual 3-D eddy current field induced. In this work, we aimed to predict, using linear response theory, the residual 3-D eddy current field required for geometric distortion correction based on phantom eddy current field measurements. The predicted 3-D eddy current field induced by the diffusion-weighting gradients was able to reduce the root mean square error of the residual eddy current field to ~1 Hz. The model's performance was tested on diffusion weighted images of four normal volunteers, following distortion correction, the quality of the Stejskal-Tanner diffusion-weighted images was found to have comparable quality to image registration based corrections (FSL) at low b-values. Unlike registration techniques the correction was not hindered by low SNR at high b-values, and results in improved image quality relative to FSL. Characterization of the 3-D eddy current field with linear response theory enables the prediction of the 3-D eddy current field required to correct eddy current induced geometric distortions for a wide range of clinical and high b-value protocols.

Mesh:

Year:  2013        PMID: 23674437     DOI: 10.1109/TMI.2013.2259249

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  3 in total

1.  Efficacy of distortion correction on diffusion imaging: comparison of FSL eddy and eddy_correct using 30 and 60 directions diffusion encoding.

Authors:  Haruyasu Yamada; Osamu Abe; Takashi Shizukuishi; Junko Kikuta; Takahiro Shinozaki; Ko Dezawa; Akira Nagano; Masayuki Matsuda; Hiroki Haradome; Yoshiki Imamura
Journal:  PLoS One       Date:  2014-11-18       Impact factor: 3.240

2.  Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging.

Authors:  Trong-Kha Truong; Allen W Song; Nan-Kuei Chen
Journal:  Biomed Res Int       Date:  2015-08-30       Impact factor: 3.411

3.  High-resolution diffusion tensor imaging of the human kidneys using a free-breathing, multi-slice, targeted field of view approach.

Authors:  Rachel W Chan; Constantin Von Deuster; Christian T Stoeck; Jack Harmer; Shonit Punwani; Navin Ramachandran; Sebastian Kozerke; David Atkinson
Journal:  NMR Biomed       Date:  2014-09-15       Impact factor: 4.044