Excessive negative venous line pressures and increased arterial air bubble counts during miniaturized cardiopulmonary bypass: an experimental study comparing miniaturized with conventional perfusion systems.

Anas Aboud1, Kai Liebing, Jochen Börgermann, Stephan Ensminger, Armin Zittermann, Andre Renner, Kavous Hakim-Meibodi, Jan Gummert.   

Abstract

OBJECTIVES: Miniaturized cardiopulmonary bypass (MCPB) is increasingly used in cardiac surgery, because it can lower clinically significant complications such as systemic inflammatory response, haemolysis and high transfusion requirements. A limitation of MCPB is the risk of excessive negative pressure in the venous line during volume depletion, probably leading to gaseous microembolism.
METHODS: In an experimental study with 24 pigs, we compared conventional open cardiopulmonary bypass (CCPB group, n = 11) with MCPB (n = 13). The same pump and identical tubing materials were used in both groups. Primary endpoints were pressure values in the venous line and the right atrium as well as the amount of air bubbles >500 µm. Secondary endpoints were biochemical parameters of systemic inflammatory response, ischaemia, haemodilution and haemolysis.
RESULTS: Nearly 20% of venous pressure values were below -150 mmHg and approximately 10% of the right atrial pressure values were below -100 mmHg in the MCPB group, during the experiment. No such low values were observed in the CCPB group. In addition, the number of large arterial air bubbles was higher in the MCPB group compared with the CCPB group (mean ± standard deviation [SD]: 13 444 ± 5709 vs 0.9 ± 0.6, respectively; P < 0.001). Bubble volume was also significantly larger during MCPB compared with CCPB (mean ± SD: 1522 ± 654 vs 4.1 ± 2.5 µl, respectively; P < 0.001). Blood levels of interleukin-6, free haemoglobin and creatine kinase were significantly higher in the CCPB group compared with the MCPB group.
CONCLUSIONS: Despite the benefits of MCPB regarding systemic inflammatory response and haemolysis, this technique is associated with excessive negative venous line pressures and a significant increase in the number and volume of arterial air bubbles compared with CCPB. Mini-perfusion systems and the management of MCPB require further refinements to avoid such adverse effects.

Entities:  

Keywords:  Cardiopulmonary bypass; Gaseous microembolism; Haemodilution; Haemolysis; Systemic inflammatory response

Mesh:

Substances:

Year:  2013        PMID: 23666376     DOI: 10.1093/ejcts/ezt257

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  2 in total

Review 1.  Use of minimal invasive extracorporeal circulation in cardiac surgery: principles, definitions and potential benefits. A position paper from the Minimal invasive Extra-Corporeal Technologies international Society (MiECTiS).

Authors:  Kyriakos Anastasiadis; John Murkin; Polychronis Antonitsis; Adrian Bauer; Marco Ranucci; Erich Gygax; Jan Schaarschmidt; Yves Fromes; Alois Philipp; Balthasar Eberle; Prakash Punjabi; Helena Argiriadou; Alexander Kadner; Hansjoerg Jenni; Guenter Albrecht; Wim van Boven; Andreas Liebold; Fillip de Somer; Harald Hausmann; Apostolos Deliopoulos; Aschraf El-Essawi; Valerio Mazzei; Fausto Biancari; Adam Fernandez; Patrick Weerwind; Thomas Puehler; Cyril Serrick; Frans Waanders; Serdar Gunaydin; Sunil Ohri; Jan Gummert; Gianni Angelini; Volkmar Falk; Thierry Carrel
Journal:  Interact Cardiovasc Thorac Surg       Date:  2016-01-26

2.  Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems.

Authors:  Anas Aboud; Hendrikje Mederos-Dahms; Kai Liebing; Armin Zittermann; Harald Schubert; Edward Murray; Andre Renner; Jan Gummert; Jochen Börgermann
Journal:  BMC Anesthesiol       Date:  2015-05-29       Impact factor: 2.217

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.