The use of biodegradable sutures for the fixation of tibial eminence fractures in children: a comparison using PDS II, Vicryl and FiberWire.

Johannes Schneppendahl1, Simon Thelen, Sören Twehues, Christian Eichler, Marcel Betsch, Joachim Windolf, Mohssen Hakimi, Michael Wild.   

Abstract

BACKGROUND: Arthroscopic suture fixation of tibial eminence fractures using FiberWire is a favorable therapeutic option. The application of biodegradable material may be of advantage especially during childhood. The aim of this study was to evaluate the biomechanical properties using the biodegradable suture materials PDS II and Vicryl compared with FiberWire.
METHODS: Bone mineral density was evaluated by pqCT in 18 human knee specimens and 3 similar groups were formed. A standardized anterior tibial eminence fracture was created and suture fixation was performed using each suture material (PDS II, Vicryl, FiberWire) in 6 specimens. Cyclic testing and destructive loading to failure was conducted.
RESULTS: Both testing modalities showed PDS II to be inferior to the other evaluated suture materials. Although significantly higher failure loads were seen with FiberWire sutures, Vicryl yielded comparable stiffness in load-to-failure testing. No significant differences between Vicryl and FiberWire could be observed under nondestructive cyclic conditions.
CONCLUSIONS: Even though FiberWire yields a superior ultimate failure load, Vicryl presented comparable results under cyclic conditions. CLINICAL RELEVANCE: For tibial eminence fractures in children, Vicryl should be considered as an alternative biodegradable suture material. The use of PDS II cannot be advocated.

Entities:  

Mesh:

Year:  2013        PMID: 23653031     DOI: 10.1097/BPO.0b013e31827d0c67

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  5 in total

1.  Absorbable and non-absorbable suture fixation results in similar outcomes for tibial eminence fractures in children and adolescents.

Authors:  Stefan Brunner; Patrick Vavken; Robert Kilger; Julia Vavken; Erich Rutz; Reinald Brunner; Carlo Camathias
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-10-31       Impact factor: 4.342

2.  Successful stabilisation of a type III paediatric tibial eminence fracture using a tensioned wire technique.

Authors:  Matthew Archer; Tom Parkin; Mark David Latimer
Journal:  BMJ Case Rep       Date:  2016-09-19

3.  Bucket handle meniscus tears in low-resource settings can be successfully treated with a cost-effective technique.

Authors:  Allicia Ostoposides Imada; James J O'Hara; Ignacio L Proumen; Pablo S Molinari; Daniel C Wascher; Dustin L Richter; Robert C Schenck
Journal:  Int Orthop       Date:  2021-05-29       Impact factor: 3.075

4.  Preliminary clinical outcomes of the double-row anchor suture-bridge technique for the fixation of tibial intercondylar eminence fractures in adults: a 12-months minimal follow-up.

Authors:  Yupeng Chu; Ting Hu; Mangmang Chen; Chendi Jiang; Zhuqi Wu; Junwu Shi
Journal:  BMC Musculoskelet Disord       Date:  2021-01-13       Impact factor: 2.362

5.  The most economical arthroscopic suture fixation for tibial intercondylar eminence avulsion fracture without any implant.

Authors:  Libo Yuan; Rongmao Shi; Zhian Chen; Wei Ding; Hongbo Tan
Journal:  J Orthop Surg Res       Date:  2022-06-25       Impact factor: 2.677

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.