Fast digital lock-in amplifier for dynamic spectrum extraction.

Mei Zhou1, Gang Li, Ling Lin.   

Abstract

An appropriate method for spectrum extraction for better signal-to-noise ratio (SNR) and lower computational cost is essential in noninvasive detection. We have first studied two existing extraction methods for dynamic spectrum (DS) comparatively: frequency domain analysis and single trial estimation; after analyzing the advantages and disadvantages theoretically, a new method based on a fast digital lock-in amplifier (FDLIA) was developed to overcome the limitations of these two existing methods. The feasibility of the new method was verified by experiments and the results demonstrated that the FDLIA method based on DS had greatly simplified the computation of frequency domain analysis without the method error; moreover, the continuous signal was cut into several short segments in FDLIA and the gross errors from an episode of pulse wave were eliminated, and thus SNR improved. Therefore, the FDLIA method utilizing the advantages of both existing methods can be effectively realized in a general embedded system in real time for its simple algorithm.

Mesh:

Substances:

Year:  2013        PMID: 23652345     DOI: 10.1117/1.JBO.18.5.057003

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  1 in total

1.  Non-linearity correction in NIR absorption spectra by grouping modeling according to the content of analyte.

Authors:  Ai Liu; Gang Li; Zhigang Fu; Yang Guan; Ling Lin
Journal:  Sci Rep       Date:  2018-06-04       Impact factor: 4.379