Literature DB >> 23645055

The clinical spectrum of longitudinal deformation of coronary stents: from a mere angiographic finding to a severe complication.

Stefano Rigattieri1, Alessandro Sciahbasi, Paolo Loschiavo.   

Abstract

BACKGROUND: Longitudinal deformation of coronary stents has been recently described and seems to be more frequent with certain contemporary stent platforms. Indeed, in order to increase flexibility and deliverability, stent manufacturers have reduced strut thickness and the number of connectors within cells; this could negatively affect other mechanical properties of the device, such as the resistance to longitudinal stress. Moreover, longitudinal deformation has been associated to adverse events, such as stent thrombosis.
METHODS: We report 3 cases of longitudinal stent deformation observed at our institution.
RESULTS: The first case was a consequence of postdilatation of the stent with a non-compliant balloon, whereas the other 2 cases involved the treatment of bifurcation lesions. One case was complicated by acute, intraprocedural stent thrombosis; such a dreadful complication, to the best of our knowledge, has not been previously reported.
CONCLUSIONS: Although longitudinal stent deformation is an infrequent finding, usually not associated with adverse events, at least in the short term, it can sometimes turn into a catastrophic, life-threatening complication. The growing number of reports about this issue in recent years should prompt the operators to carefully select coronary stents, especially when dealing with certain lesion subsets, such as ostial lesions, bifurcations, and long lesions.

Entities:  

Mesh:

Year:  2013        PMID: 23645055

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  3 in total

1.  Coronary artery perforation after crush technique for stent compression.

Authors:  A Avci; S Fidan; G Alıcı; G Acar; C Toprak; S Izci; A M Esen
Journal:  Herz       Date:  2014-04-17       Impact factor: 1.443

2.  Stent accordion phenomenon visualized with optical coherence tomography.

Authors:  Lucky R Cuenza; Khung Keong Yeo
Journal:  Cardiovasc Diagn Ther       Date:  2017-12

3.  Longitudinal Stent Deformation at Aneurysm Site: Flexibility at the Expense of Longitudinal Integrity.

Authors:  Bhupesh R Shah
Journal:  J Clin Diagn Res       Date:  2017-07-01
  3 in total