Literature DB >> 23642342

Prediction of seizure likelihood with a long-term, implanted seizure advisory system in patients with drug-resistant epilepsy: a first-in-man study.

Mark J Cook1, Terence J O'Brien, Samuel F Berkovic, Michael Murphy, Andrew Morokoff, Gavin Fabinyi, Wendyl D'Souza, Raju Yerra, John Archer, Lucas Litewka, Sean Hosking, Paul Lightfoot, Vanessa Ruedebusch, W Douglas Sheffield, David Snyder, Kent Leyde, David Himes.   

Abstract

BACKGROUND: Seizure prediction would be clinically useful in patients with epilepsy and could improve safety, increase independence, and allow acute treatment. We did a multicentre clinical feasibility study to assess the safety and efficacy of a long-term implanted seizure advisory system designed to predict seizure likelihood and quantify seizures in adults with drug-resistant focal seizures.
METHODS: We enrolled patients at three centres in Melbourne, Australia, between March 24, 2010, and June 21, 2011. Eligible patients had between two and 12 disabling partial-onset seizures per month, a lateralised epileptogenic zone, and no history of psychogenic seizures. After devices were surgically implanted, patients entered a data collection phase, during which an algorithm for identification of periods of high, moderate, and low seizure likelihood was established. If the algorithm met performance criteria (ie, sensitivity of high-likelihood warnings greater than 65% and performance better than expected through chance prediction of randomly occurring events), patients then entered an advisory phase and received information about seizure likelihood. The primary endpoint was the number of device-related adverse events at 4 months after implantation. Our secondary endpoints were algorithm performance at the end of the data collection phase, clinical effectiveness (measures of anxiety, depression, seizure severity, and quality of life) 4 months after initiation of the advisory phase, and longer-term adverse events. This trial is registered with ClinicalTrials.gov, number NCT01043406.
FINDINGS: We implanted 15 patients with the advisory system. 11 device-related adverse events were noted within four months of implantation, two of which were serious (device migration, seroma); an additional two serious adverse events occurred during the first year after implantation (device-related infection, device site reaction), but were resolved without further complication. The device met enabling criteria in 11 patients upon completion of the data collection phase, with high likelihood performance estimate sensitivities ranging from 65% to 100%. Three patients' algorithms did not meet performance criteria and one patient required device removal because of an adverse event before sufficient training data were acquired. We detected no significant changes in clinical effectiveness measures between baseline and 4 months after implantation.
INTERPRETATION: This study showed that intracranial electroencephalographic monitoring is feasible in ambulatory patients with drug-resistant epilepsy. If these findings are replicated in larger, longer studies, accurate definition of preictal electrical activity might improve understanding of seizure generation and eventually lead to new management strategies. FUNDING: NeuroVista.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23642342     DOI: 10.1016/S1474-4422(13)70075-9

Source DB:  PubMed          Journal:  Lancet Neurol        ISSN: 1474-4422            Impact factor:   44.182


  179 in total

1.  Minimally invasive endovascular stent-electrode array for high-fidelity, chronic recordings of cortical neural activity.

Authors:  Thomas J Oxley; Nicholas L Opie; Sam E John; Gil S Rind; Stephen M Ronayne; Tracey L Wheeler; Jack W Judy; Alan J McDonald; Anthony Dornom; Timothy J H Lovell; Christopher Steward; David J Garrett; Bradford A Moffat; Elaine H Lui; Nawaf Yassi; Bruce C V Campbell; Yan T Wong; Kate E Fox; Ewan S Nurse; Iwan E Bennett; Sébastien H Bauquier; Kishan A Liyanage; Nicole R van der Nagel; Piero Perucca; Arman Ahnood; Katherine P Gill; Bernard Yan; Leonid Churilov; Christopher R French; Patricia M Desmond; Malcolm K Horne; Lynette Kiers; Steven Prawer; Stephen M Davis; Anthony N Burkitt; Peter J Mitchell; David B Grayden; Clive N May; Terence J O'Brien
Journal:  Nat Biotechnol       Date:  2016-02-08       Impact factor: 54.908

2.  Bursts of seizures in long-term recordings of human focal epilepsy.

Authors:  Philippa J Karoly; Ewan S Nurse; Dean R Freestone; Hoameng Ung; Mark J Cook; Ray Boston
Journal:  Epilepsia       Date:  2017-01-13       Impact factor: 5.864

3.  Predicting seizures: are we there yet?

Authors:  Sheryl Haut
Journal:  Epilepsy Curr       Date:  2013-11       Impact factor: 7.500

Review 4.  Deep Brain Stimulation for Epilepsy: Biomarkers for Optimization.

Authors:  Katrina L Dell; Mark J Cook; Matias I Maturana
Journal:  Curr Treat Options Neurol       Date:  2019-09-26       Impact factor: 3.598

Review 5.  Seizure detection: do current devices work? And when can they be useful?

Authors:  Xiuhe Zhao; Samden D Lhatoo
Journal:  Curr Neurol Neurosci Rep       Date:  2018-05-23       Impact factor: 5.081

6.  Rapid eye movement sleep and hippocampal theta oscillations precede seizure onset in the tetanus toxin model of temporal lobe epilepsy.

Authors:  Madineh Sedigh-Sarvestani; Godfrey I Thuku; Sridhar Sunderam; Anjum Parkar; Steven L Weinstein; Steven J Schiff; Bruce J Gluckman
Journal:  J Neurosci       Date:  2014-01-22       Impact factor: 6.167

7.  Cloud computing for seizure detection in implanted neural devices.

Authors:  Steven Baldassano; Xuelong Zhao; Benjamin Brinkmann; Vaclav Kremen; John Bernabei; Mark Cook; Timothy Denison; Gregory Worrell; Brian Litt
Journal:  J Neural Eng       Date:  2018-12-18       Impact factor: 5.379

Review 8.  Physiological properties of brain-machine interface input signals.

Authors:  Marc W Slutzky; Robert D Flint
Journal:  J Neurophysiol       Date:  2017-06-14       Impact factor: 2.714

9.  Intracranial EEG fluctuates over months after implanting electrodes in human brain.

Authors:  Hoameng Ung; Steven N Baldassano; Hank Bink; Abba M Krieger; Shawniqua Williams; Flavia Vitale; Chengyuan Wu; Dean Freestone; Ewan Nurse; Kent Leyde; Kathryn A Davis; Mark Cook; Brian Litt
Journal:  J Neural Eng       Date:  2017-09-01       Impact factor: 5.379

Review 10.  Interictal high-frequency oscillations in focal human epilepsy.

Authors:  Jan Cimbalnik; Michal T Kucewicz; Greg Worrell
Journal:  Curr Opin Neurol       Date:  2016-04       Impact factor: 5.710

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.