Correlation between cataract grading by Scheimpflug imaging and phaco time in phacoemulsification using peristaltic and venturi pumps.

Maria Sara Patrício1, Ana Catarina Almeida, Maria Picoto Rodrigues, Marta Esteves Guedes, Tiago Bravo Ferreira.   

Abstract

PURPOSE: To study the correlation among cataract grading system by Scheimpflug imaging (Oculus Pentacam® Nucleus Grading System [PNS]), effective phaco time (EPT), and surgical time (ST) using peristaltic and venturi pumps on the Whitestar Signature Phaco System (Abbott Medical Optics, Inc., Santa Ana, California, USA).
METHODS: Patients scheduled to have bimanual microincision phacoemulsification by the same surgeon (T.F.) had Pentacam® Scheimpflug assessment using PNS and were randomly distributed into 
2 groups. In group 1, surgery was performed using peristaltic pump, and in group 2, using venturi pump. Parameters assessed in both groups included EPT and ST.
RESULTS: Sixty eyes that had uneventful clear cornea phacoemulsification with implantation of foldable intraocular lens (IOL) were evaluated, 30 in each group. The EPT was statistically superior with higher nuclear density grades for both groups 1 and 2 (group 1 Spearman R = 0.739, p<0.001; group 2 Spearman R = 0.924, p<0.001). The ST statistically increased with nucleus density for both groups 1 and 2 (group 
1 Spearman R = 0.620, p<0.001; group 2 Spearman R = 0.643, p<0.001). Group 2 presented statistically significant reduction on median ST compared with group 1 for nucleus grading 2, 3, 4, and 5.
CONCLUSIONS: The PNS is an objective and reproducible method of grading nuclear cataract density and had a correlation with EPT. Peristaltic and venturi pumps in Signature Phaco System were similar in EPT but venturi pump reduced the median ST in nucleus with higher PNS, without repercussion on complications or anterior chamber stability.

Entities:  

Mesh:

Year:  2013        PMID: 23640507     DOI: 10.5301/ejo.5000293

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  1 in total

1.  Prospective randomized comparative study between venturi and peristaltic pumps in WhiteStar Signature® phacoemulsification machine.

Authors:  Wilson Takashi Hida; André Lins de Medeiros; André Gustavo de Araújo Rolim; Antonio Francisco Pimenta Motta; Danilo Varela Kniggendorf; Rodrigo Lafetá Franco de Queiroz; Mário Augusto Pereira Dias Chaves; Pedro Carlos Carricondo; Celso Takashi Nakano; Walton Nosé
Journal:  Clin Ophthalmol       Date:  2018-12-27
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.