A comparison of applanation tonometry using conventional reusable goldmann prisms and disposable prisms.

Andrew S H Tsai1, Alicia C How, Daniel H W Su, Kelvin Y Lee, Tina T Wong, Shamira A Perera.   

Abstract

PURPOSE: To determine the agreement between intraocular pressure (IOP) measurements using conventional Goldmann applanation tonometry (GAT) and Tonosafe disposable prisms, and also to provide a comprehensive cost analysis of the use of both types of prisms
METHODS: : In this prospective observational study, 198 eyes of 100 glaucoma patients had their IOPs measured by 5 consultant ophthalmologists. Data were analyzed using the Bland-Altman method of differences, and correlation was measured using the Pearson coefficient. An analysis of the cost incurred using the 2 methods over a 6-month period was performed.
RESULTS: The majority were Chinese (82%), with a male preponderance (57%). The range of IOPs as measured by GAT was 4 to 34 mm Hg. Using the Bland-Altman method to compare GAT and disposable prisms, the bias was 0.2 mm Hg. Tonosafe overestimated the IOP by 0.2 mm Hg in the right eye and underestimated it by 0.2 mm Hg in the left eye. The Tonosafe IOP correlated well with GAT, with a Pearson coefficient of correlation(r) of 0.91 (P<0.0005) for the right eye and 0.92 (P<0.0005) for the left eye, respectively. For those with GAT IOP≥21 mm Hg (n=26), Tonosafe underestimated the IOP by 0.35 mm Hg. The cost incurred by Tonosafe prisms was approximately 8 times that of GAT, but the cost differential reverses when GAT had to be replaced after every 100 cycles of disinfection.
CONCLUSIONS: We found a good correlation between Tonosafe prisms and conventional GAT in measuring the IOP. Tonosafe prisms may be of use, especially if the risk of transmission of infection is high. However, cost may limit its more widespread use.

Entities:  

Mesh:

Year:  2014        PMID: 23632407     DOI: 10.1097/IJG.0b013e318294863d

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  3 in total

Review 1.  How should we measure intraocular pressure in the era of coronavirus disease 2019? Balancing infectious risk, cleaning requirements, and accuracy.

Authors:  Christine A Petersen; Andrew Chen; Philip P Chen
Journal:  Curr Opin Ophthalmol       Date:  2022-03-01       Impact factor: 3.761

Review 2.  Guidelines and Recommendations for Tonometry Use during the COVID-19 Era.

Authors:  Enmar M Almazyad; Sally Ameen; Mohammad A Khan; Rizwan Malik
Journal:  Middle East Afr J Ophthalmol       Date:  2020-07-20

Review 3.  How glaucoma care changed for the better after the pandemic.

Authors:  Kateki Vinod; Paul A Sidoti
Journal:  Curr Opin Ophthalmol       Date:  2022-03-01       Impact factor: 3.761

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.