A portable device for measuring donor corneal transparency in eye banks.

Mohit Parekh1, Stefano Ferrari, Alessandro Ruzza, Mariarosaria Pugliese, Diego Ponzin, Gianni Salvalaio.   

Abstract

To develop a portable device for measuring the donor corneal transparency and validate its efficacy for corneal evaluation in the eye-banks and for research. The transparency device (TD) has a light source, a detachable system for corneal insertion and a base for light transmission. The probe detects the transmitted light which is measured by a lux-meter. A contact lens was set as 'control' to reduce the light scattering concern, an empty petri-plate as 'blank' and the cornea as 'sample'. Two experts and non-experts (masked) observed the corneas for subjective analysis which was then compared using the TD. The parameters observed were scars, foreign-body, stromal-deformities, folds, thickness and opacity which were then converted to a relative overall percentage by the observer. Twenty corneas were evaluated for correlation, five tissues to obtain standard-deviation and twenty-four pairs for a comparative study. Experts mimicked the eye-banks with long-term experience while non-experts mimicked the emerging eye-banks. Subjective values by the experts closely resembled the measurements by TD. The average correlation between the experts and the non-experts to TD was 0.985 and 0.960 respectively. TD showed higher reproducibility than experts followed by the non-experts. The comparative study showed that increase in thickness reduces the transparency. TD is portable, easy, efficient, maintains sterility and less expensive hence the emerging eye-banks and researchers can use to raise their standards and evaluate the transparency for in vitro tests and comparative studies. The suitable transparency for the cornea deemed for clinical applications was found to be >75 %.

Entities:  

Mesh:

Year:  2013        PMID: 23625054     DOI: 10.1007/s10561-013-9372-2

Source DB:  PubMed          Journal:  Cell Tissue Bank        ISSN: 1389-9333            Impact factor:   1.522


  2 in total

Review 1.  Corneal structure and transparency.

Authors:  Keith M Meek; Carlo Knupp
Journal:  Prog Retin Eye Res       Date:  2015-07-02       Impact factor: 21.198

2.  Incremental Concentrations of Tacrolimus Eye Drops as a Strategy for the Management of Severe Vernal Keratoconjunctivitis.

Authors:  Maan Abdullah Albarry; Mohit Parekh; Stefano Ferrari; Heba Mahmoud Eltahir; Ahmed M Shehata; Mohamed A Shaker; Hossein Mostafa Elbadawy
Journal:  Front Pharmacol       Date:  2022-03-25       Impact factor: 5.810

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.