Literature DB >> 23623494

Effect of an orthosis on the loads acting on a vertebral body replacement.

Antonius Rohlmann1, Thomas Zander, Friedmar Graichen, Georg Bergmann.   

Abstract

BACKGROUND: The spinal load reduction by an orthosis is still a matter of debate. Some studies predicted a load reduction while others found no effect. The aim of this study was to measure the in vivo effect of the Lumbo TriStep brace and the hyperextension orthosis medi 3C on the spinal implant loads.
METHODS: Telemeterized vertebral body replacements were implanted in 5 patients suffering from a severe fracture of the L1 or L3 vertebral body. The implant allows the measurement of 6 load components acting on it. For several activities during standing, sitting and walking, implant loads were measured in patients with and without an orthosis.
FINDINGS: The average resultant force on the vertebral body for 26 activities was reduced by 9% with the Lumbo TriStep brace, and by 19% with the hyperextension orthosis. The force reduction is usually more pronounced for activities performed during sitting than it is for those performed while standing. However, considerable inter- and intra-individual variation was observed. In several cases, the measured implant forces were even higher when the patients were wearing an orthosis.
INTERPRETATION: In some patients, for certain activities, an orthosis may reduce the force on a vertebral body replacement and thus on the anterior column of the spine. However, in other patients for the same activities, an orthosis may increase the force. The measurements do not allow a clear recommendation to wear an orthosis since the clinically relevant reduction of implant forces is unknown.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23623494     DOI: 10.1016/j.clinbiomech.2013.03.010

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

Review 1.  'SMART' implantable devices for spinal implants: a systematic review on current and future trends.

Authors:  Sihyong J Kim; Tian Wang; Matthew H Pelletier; William R Walsh
Journal:  J Spine Surg       Date:  2022-03

2.  Spinal loads during cycling on an ergometer.

Authors:  Antonius Rohlmann; Thomas Zander; Friedmar Graichen; Hendrik Schmidt; Georg Bergmann
Journal:  PLoS One       Date:  2014-04-17       Impact factor: 3.240

3.  Is postoperative bracing after pedicle screw fixation of spine fractures necessary? Study protocol of the ORNOT study: a randomised controlled multicentre trial.

Authors:  Arjen Johannes Smits; Jaap Deunk; Agnita Stadhouder; Mark Cornelis Altena; Diederik Hendrik Ruth Kempen; Frank Willem Bloemers