Light curing in orthodontics; should we be concerned?

Neil McCusker1, Siu Man Lee, Stephen Robinson, Naresh Patel, Jonathan R Sandy, Anthony J Ireland.   

Abstract

OBJECTIVES: Light cured materials are increasingly used in orthodontic clinical practice and concurrent with developments in materials have been developments in light curing unit technology. In recent years the irradiances of these units have increased. The aim of this study was to determine the safe exposure times to both direct and reflected light.
METHODS: The weighted irradiance and safe exposure times of 11 dental curing lights (1 plasma arc, 2 halogen and 8 LED lights) were determined at 6 distances (2-60 cm) from the light guide tip using a spectroradiometer. In addition, using the single most powerful light, the same two parameters were determined for reflected light. This was done at a distance of 10 cm from the reflected light, but during simulated bonding of 8 different orthodontic brackets of three material types, namely stainless steel, ceramic and composite.
RESULTS: The results indicate that the LED Fusion lamp had the highest weighted irradiance and the shortest safe exposure time. With this light the maximum safe exposure time without additional eye protection for the patient (at 10 cm), the operator (at 30 cm) and the assistant (at 60 cm) ranged from 2.5 min, 22.1 min and 88.8 min respectively. This indicates a relatively low short term risk during normal operation of dental curing lights. For reflected light at a distance of 10 cm the risk was even lower, but was affected by the material and shape of the orthodontic bracket under test. SIGNIFICANCE: The short term risks associated with the use of dental curing lights, halogen, LED or plasma, appear to be low, particularly if as is the case adequate safety precautions are employed. The same is true for reflected light from orthodontic brackets during bonding. What is still unclear is the potential long term ocular effects of prolonged exposure to the blue light generated from dental curing lights.
Copyright © 2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23618555     DOI: 10.1016/j.dental.2013.03.023

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  3 in total

1.  Guidelines for the selection, use, and maintenance of LED light-curing units - Part II.

Authors:  A C Shortall; R B Price; L MacKenzie; F J T Burke
Journal:  Br Dent J       Date:  2016-11-04       Impact factor: 1.626

Review 2.  Ocular hazards of curing light units used in dental practice - A systematic review.

Authors:  Raniya A Alasiri; Hashim A Algarni; Reem A Alasiri
Journal:  Saudi Dent J       Date:  2019-02-20

3.  Non-Ionizing Radiation in Swedish Health Care-Exposure and Safety Aspects.

Authors:  Kjell Hansson Mild; Ronnie Lundström; Jonna Wilén
Journal:  Int J Environ Res Public Health       Date:  2019-04-02       Impact factor: 3.390

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.