Literature DB >> 23615097

Safety of hybrid electrodes for single-neuron recordings in humans.

Stefan Hefft1, Armin Brandt, Stefan Zwick, Dominik von Elverfeldt, Irina Mader, Joacir Cordeiro, Michael Trippel, Julie Blumberg, Andreas Schulze-Bonhage.   

Abstract

BACKGROUND: Intracranial in vivo recordings of individual neurons in humans are increasingly performed for a better understanding of the mechanisms of epileptogenesis and of the neurobiological basis of cognition. So far, information about the safety of stereotactic implantations and of magnetic resonance imaging (MRI) with hybrid depth electrodes is scarce.
OBJECTIVE: The aim of this study was to assess neurosurgical safety of implantations, recordings, and imaging using hybrid electrodes in humans.
METHODS: Perioperative and long-term safety of implantation of a total of 88 hybrid depth electrodes with integrated microwires was assessed retrospectively in 25 consecutive epilepsy patients who underwent implantation of electrodes from 2007 to 2011 based on electronically stored charts. Safety aspects of MRI are reported from both in vitro and in vivo investigations. Precision of electrode implantation is evaluated based on intraoperative computed tomography and pre- and postoperative MRI.
RESULTS: There was no clinically relevant morbidity associated with the use of hybrid electrodes in any of the patients. Precision of recordings from the targets aimed at was similar to that of standard depth electrodes. In vitro studies demonstrated the absence of relevant heating of hybrid electrodes with newly designed connectors with MRI at 1.5 T, corresponding to well-tolerated clinical MRI in patients.
CONCLUSION: Given the technical approach described here, precise targeting and safe use are possible with hybrid electrodes containing microwires for in vivo recording of human neuronal units.

Entities:  

Mesh:

Year:  2013        PMID: 23615097     DOI: 10.1227/01.neu.0000429840.76460.8c

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  6 in total

Review 1.  The preoperative evaluation and surgical treatment of epilepsy.

Authors:  Andreas Schulze-Bonhage; Josef Zentner
Journal:  Dtsch Arztebl Int       Date:  2014-05-02       Impact factor: 5.594

2.  Safety and Utility of Hybrid Depth Electrodes for Seizure Localization and Single-Unit Neuronal Recording.

Authors:  April A Carlson; Ueli Rutishauser; Adam N Mamelak
Journal:  Stereotact Funct Neurosurg       Date:  2018-10-16       Impact factor: 1.875

3.  Interictal spikes with and without high-frequency oscillation have different single-neuron correlates.

Authors:  Tim A Guth; Lukas Kunz; Armin Brandt; Matthias Dümpelmann; Kerstin A Klotz; Peter C Reinacher; Andreas Schulze-Bonhage; Julia Jacobs; Jan Schönberger
Journal:  Brain       Date:  2021-11-29       Impact factor: 15.255

Review 4.  Stimulation Mapping Using Stereoelectroencephalography: Current and Future Directions.

Authors:  Derek D George; Steven G Ojemann; Cornelia Drees; John A Thompson
Journal:  Front Neurol       Date:  2020-05-12       Impact factor: 4.003

Review 5.  Poly(3,4-ethylenedioxythiophene)-Based Neural Interfaces for Recording and Stimulation: Fundamental Aspects and In Vivo Applications.

Authors:  Michele Bianchi; Anna De Salvo; Maria Asplund; Stefano Carli; Michele Di Lauro; Andreas Schulze-Bonhage; Thomas Stieglitz; Luciano Fadiga; Fabio Biscarini
Journal:  Adv Sci (Weinh)       Date:  2022-02-21       Impact factor: 17.521

Review 6.  How Human Single-Neuron Recordings Can Help Us Understand Cognition: Insights from Memory Studies.

Authors:  Zuzanna Roma Kubska; Jan Kamiński
Journal:  Brain Sci       Date:  2021-03-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.