Literature DB >> 23612554

Automated indirect immunofluorescence antinuclear antibody analysis is a standardized alternative for visual microscope interpretation.

Carolien Bonroy, Charlotte Verfaillie, Vanessa Smith, Lies Persijn, Evy De Witte, Filip De Keyser, Katrien Devreese.   

Abstract

BACKGROUND: Screening for antinuclear antibodies (ANA) is a basic tool in the serological work-up of systemic rheumatic disorders. Despite the emergence of alternative screening methods and the difficulties in standardization, indirect immunofluorescence (IIF) remains the recommended method for ANA detection. This study aimed to assess the reliability of automated ANA IIF analysis as a standardized alternative for the conventional manual approach.
METHODS: ANA testing on HEp-2000 cells was performed on 304 consecutive routine sera, 28 serumbank samples displaying rare staining patterns, 219 samples of well-defined disease cohorts [141 systemic sclerosis (SSc), 13 polymyalgia rheumatica, 22 osteoarthritis, 5 ANCA-associated vasculitis and 38 spondyloarthritis] and 100 healthy donors. All samples were analyzed by automated IIF (Zenit G-sight), by conventional visual IIF microscopy and two ANA screening enzyme immunoassays (EIA).
RESULTS: Automated and conventional ANA IIF analysis were comparable for negative/positive interpretation as well as intensity assessment (>90% agreement). In contrast, the accuracy of pattern recognition (26%) was limited. Likelihood ratios (LR) for SSc on results intervals of both Zenit G-sight and EIA increased with increasing level of positivity. Sensitivity within the SSc-associated antibody subsets was higher for Zenit G-sight (97%-100%) than EIA (10%-96%). A significant correlation between the quantitative result obtained by Zenit G-sight and the conventional end-point titer was found.
CONCLUSIONS: The use of Zenit G-sight for automated ANA IIF analysis offers opportunities to improve standardization. However, a complementary role of the expert technicians remains, especially for pattern recognition and classification of uncertain/negative samples.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23612554     DOI: 10.1515/cclm-2013-0016

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  18 in total

1.  A fully automated IIF system for the detection of antinuclear antibodies and antineutrophil cytoplasmic antibodies.

Authors:  O Shovman; N Agmon-Levin; B Gilburd; T Martins; A Petzold; T Matthias; Y Shoenfeld
Journal:  Immunol Res       Date:  2015-02       Impact factor: 2.829

2.  Clinical performance evaluation of a novel, automated chemiluminescent immunoassay, QUANTA Flash CTD Screen Plus.

Authors:  Chelsea Bentow; Gabriella Lakos; Rachel Rosenblum; Cassandra Bryant; Andrea Seaman; Michael Mahler
Journal:  Immunol Res       Date:  2015-02       Impact factor: 2.829

3.  The burden of the variability introduced by the HEp-2 assay kit and the CAD system in ANA indirect immunofluorescence test.

Authors:  M Infantino; F Meacci; V Grossi; M Manfredi; M Benucci; M Merone; P Soda
Journal:  Immunol Res       Date:  2017-02       Impact factor: 2.829

Review 4.  Recent Approaches To Optimize Laboratory Assessment of Antinuclear Antibodies.

Authors:  Anne E Tebo
Journal:  Clin Vaccine Immunol       Date:  2017-12-05

Review 5.  Standardization and Quality Assessment Under the Perspective of Automated Computer-Assisted HEp-2 Immunofluorescence Assay Systems.

Authors:  Luigi Cinquanta; Nicola Bizzaro; Giampaola Pesce
Journal:  Front Immunol       Date:  2021-02-25       Impact factor: 7.561

6.  Original approach for automated quantification of antinuclear autoantibodies by indirect immunofluorescence.

Authors:  Daniel Bertin; Noémie Jourde-Chiche; Pierre Bongrand; Nathalie Bardin
Journal:  Clin Dev Immunol       Date:  2013-12-18

Review 7.  Current concepts and future directions for the assessment of autoantibodies to cellular antigens referred to as anti-nuclear antibodies.

Authors:  Michael Mahler; Pier-Luigi Meroni; Xavier Bossuyt; Marvin J Fritzler
Journal:  J Immunol Res       Date:  2014-04-27       Impact factor: 4.818

8.  Interpretation of ANA indirect immunofluorescence test outside the darkroom using NOVA view compared to manual microscopy.

Authors:  Susan S Copple; Troy D Jaskowski; Rashelle Giles; Harry R Hill
Journal:  J Immunol Res       Date:  2014-02-24       Impact factor: 4.818

Review 9.  Automated tests of ANA immunofluorescence as throughput autoantibody detection technology: strengths and limitations.

Authors:  Pier Luigi Meroni; Nicola Bizzaro; Ilaria Cavazzana; Maria Orietta Borghi; Angela Tincani
Journal:  BMC Med       Date:  2014-03-03       Impact factor: 8.775

Review 10.  Novel clinical and diagnostic aspects of antineutrophil cytoplasmic antibodies.

Authors:  Johannes Schulte-Pelkum; Antonella Radice; Gary L Norman; Marcos Lόpez Hoyos; Gabriella Lakos; Carol Buchner; Lucile Musset; Makoto Miyara; Laura Stinton; Michael Mahler
Journal:  J Immunol Res       Date:  2014-06-05       Impact factor: 4.818

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.