| Literature DB >> 23608406 |
J Casado1, I Rodríguez, I Carpinteiro, M Ramil, R Cela.
Abstract
In this research, a simplified procedure for the selective determination of nine benzotriazole UV stabilizers (BUVSs) in sludge from urban sewage treatment plants is presented. Analytes were extracted from the sample using the matrix solid-phase dispersion (MSPD) technique and further determined by gas chromatography (GC) with quadrupole time-of-flight mass spectrometry (QTOF-MS). Highly selective chromatographic records were attained considering a mass window of 0.005Da, centred in the quantification product ion corresponding to each compound. Moreover, the availability of accurate ion product scan MS/MS spectra permitted to confirm the identities of peaks observed in extracted ion MS/MS chromatograms. As a result, a straightforward sample preparation procedure combining extraction and clean-up in the same step, and consuming just 10mL of ethyl acetate, sufficed to deal with complex sludge samples. The developed method attained limits of quantification (LOQs) between 2ngg(-1) and 10ngg(-1), referred to freeze-dried sludge, and recoveries from 70% to 111%, with standard deviations from 2% to 13%. Analysis of sludge samples and certified reference materials confirmed the existence of residues of eight out of nine BUVSs. UV-326, UV-328 and UV-234 displayed the highest occurrence frequencies and individual concentrations above 100ngg(-1) in several samples.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23608406 DOI: 10.1016/j.chroma.2013.03.050
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759