Impact of phase of respiration on central venous catheter tip position.

Patrick P Pan1, Bjorn I Engstrom, Matthew P Lungren, Danielle M Seaman, Mark L Lessne, Charles Y Kim.   

Abstract

PURPOSE: To determine the impact of the phase of respiration on CVC tip position using cross-sectional imaging.
METHODS: We retrospectively analyzed the CT scans of 24 consecutive patients (eight men and 16 women, mean age 56.3 years, range 18-79) who underwent a CT scan protocol that includes both imaging of the thorax in inspiration and expiration. Only patients with a central venous catheter and absence of any substantial pulmonary pathology that might affect lung volumes were included. Measurements of the catheter tip location and central venous structures were obtained from inspiratory and expiratory phase images in each patient and compared using the paired <i>t</i> test.
RESULTS: The length of the SVC and superior mediastinum were significantly longer during inspiration compared to expiration (9 mm and 7 mm respectively, P<0.001 for both). The distance between the superior and inferior cavo-atrial junction did not change significantly with respiration. The catheter tip location moved on average 9 mm (range 0-25 mm) cephalad during inspiration compared to expiration (P=0.001) in relation to the superior cavoatrial junction. The amount of catheter tip movement correlated significantly with the degree of diaphragmatic excursion with respiration (R=0.58). During inspiration, the cavo-atrial junction was on average 11 mm inferior to the right cardiomediastinal angle observed on radiography, but was nearly identical during expiration (R=0.78, P<0.001).
CONCLUSIONS: The central catheter tip position varied significantly with respiratory motion, with a mean excursion of 9 mm. The right cardiomediastinal border demonstrated a strong correlation with the actual location of the superior cavo-atrial junction in expiration, but not in inspiration.

Entities:  

Mesh:

Year:  2013        PMID: 23599138     DOI: 10.5301/jva.5000135

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  3 in total

1.  Central venous catheter tip migration due to tracheal extubation: a prospective randomized study.

Authors:  Manuel F Struck; Theresa Jünemann; Konrad Reinhart; Wolfram Schummer
Journal:  J Clin Monit Comput       Date:  2016-07-28       Impact factor: 2.502

2.  Chest radiography for simplified evaluation of central venous catheter tip positioning for safe and accurate haemodynamic monitoring: a retrospective observational study.

Authors:  Minwoo Kang; Jinkun Bae; Sujin Moon; Tae Nyoung Chung
Journal:  BMJ Open       Date:  2021-01-04       Impact factor: 2.692

Review 3.  Central Venous Catheter Intravascular Malpositioning: Causes, Prevention, Diagnosis, and Correction.

Authors:  Carlos J Roldan; Linda Paniagua
Journal:  West J Emerg Med       Date:  2015-10-20