| Literature DB >> 23599134 |
Andrew A Hill1, Thodur Vasudevan, Nathaniel P Young, Mark Crawford, Duane D Blatter, Emma Marsh, Trent Perry, Doug Smith, Chris Phillips.
Abstract
PURPOSE: The aim of this work was to assess the safety and efficacy of the VWING Vascular Needle Guide to assist in cannulation of difficult or impossible to access fistulae using the buttonhole cannulation technique.Entities:
Mesh:
Year: 2013 PMID: 23599134 PMCID: PMC6159821 DOI: 10.5301/jva.5000152
Source DB: PubMed Journal: J Vasc Access ISSN: 1129-7298 Impact factor: 2.283
Fig. 1VWING: first (left) and second (right) versions.
Characteristics of Patients Undergoing Vwing Mplantation
| Characteristics | |
|---|---|
| N | 9 |
| Mean Age (y) | 60 |
| Sex | |
| Male | 2 |
| Female | 7 |
| Ethnicity | |
| European/NZ | 2 |
| Maori | 2 |
| Pacific Islander | 5 |
| Fistula Type | |
| Radio-cephalic | 3 |
| Brachio-cephalic | 4 |
| Transposed Brachio-basilic | 2 |
| Mean Fistula Age (m) | 14 |
| Reason for Inclusion | |
| Short Segment | 5 |
| Deep | 3 |
| Mobile | 1 |
| Catheter Dependent | 4 |
m = months; y = years.
Cannulation Observations
| VWING Site | Non-Implant Site | |
|---|---|---|
| Total Access Procedures | 387 | 355 |
| Successful Site Use (%) | 94 | 77 |
| Successful Cannulation (%) | 94 | 96 |
| Single Stick Success (%) | 90 | 94 |
| Insertion Ease (1: Easy, 10: Impossible) | 1. 52 (SD 0.97) | 1. 67 (SD 1.45) |
| Insertion Pain (1: none, 10: extreme) | 0.46 (SD 1.11) | 0.382 (SD 0.79) |
| Time to Hemostasis (min.) | 8.99 (SD 2.53) | 7.9 5 (SD 2.66) |
| Backwall Puncture (%) | 1.0 | 0.7 |
| Bleeding Around Needle (%) | 2.7 | 4.6 |
| Hematoma (%) | 0.6 | 0 |
| Infiltration (%) | 1.2 | 0 |
| Overall Complications (%) | 4.8 | 4.3 |
| Interventions | 0 | 4 |
Fig. 2Fistulogram showing a VWING on the Fistula. The left arrow indicates the nondevice cannulation segment. The right arrow points to the VWING device.