| Literature DB >> 23592397 |
Luca Testa1, Guido Gelpi, Francesco Bedogni.
Abstract
We hereby present the case of a patient with severe aortic stenosis who underwent in her previous medical history a mitral valve replacement with a mechanical valve (Omnicarbon 27), and progressively developed a severe aortic stenosis. This patient was judged inoperable and then scheduled for CoreValve Revalving System implantation. Despite a good positioning of the CoreValve, an acute, severe mitral regurgitation developed soon after implantation as a consequence of the impaired movement of the mitral prosthesis leaflet. A condition of cardiogenic shock quickly developed. A good mitral prosthesis function was restored disengaging the CoreValve from the aortic annulus. After few months, the patients underwent successful Edwards-Sapien valve implantation through the Corevalve. This case strongly demonstrates how much a careful evaluation of the features of the mitral prosthesis and patient anatomy is crucial to select which specific transcatheter bioprosthesis would better perform.Entities:
Keywords: complications adult cath/intervention; congestive heart failure; valvular heart disease
Mesh:
Year: 2013 PMID: 23592397 DOI: 10.1002/ccd.24948
Source DB: PubMed Journal: Catheter Cardiovasc Interv ISSN: 1522-1946 Impact factor: 2.692