| Literature DB >> 23587553 |
Xi Luo1, Ning Fang Cai, Ze Neng Cheng.
Abstract
A liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated for the determination of recombinant urate oxidase in human plasma. This assay was based on the determination of enzyme reaction product, (15)N-allantoin, and phenacetin was used as an internal standard (IS). Separation was achieved on a C18 column by the mobile phase of 30% water (containing 0.5% formic acid) and 70% methanol. Quantification was done using multiple reaction monitoring (MRM) mode to monitor the precursor-to-product ion transitions of m/z 161 → m/z 118 for (15)N-allantoin and m/z 180 → m/z 110.1 for IS at positive ionization mode. The calibration curve was established over the range of 2.077-42.06 U/l and the correlation coefficient was larger than 0.99. The intra-day and inter-day relative standard deviations were less than 10.6%. Accuracy determined at three concentrations ranged between 98.6% and 109.2%. This method was successfully applied to a pharmacokinetic study of intravenous recombinant urate oxidase produced from Escherichia coli in Chinese healthy volunteers.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23587553 DOI: 10.1016/j.jpba.2013.03.015
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935