Literature DB >> 23580122

Difficult OptEase filter retrievals after prolonged indwelling times.

Thuong G Van Ha1, Lisa Kang, Jonathan Lorenz, Steven Zangan, Rakesh Navuluri, Christopher Straus, Brian Funaki.   

Abstract

PURPOSE: The OptEase vena cave filter (Cordis, Piscataway, NJ) is commercially available as a retrievable or permanent filter with short recommended indwelling time, presumably due to extensive contact of the filter side struts with the inferior vena cava wall and subsequent neointimal hyperplasia leading to incorporation. Our purpose was to evaluate OptEase filter retrievals with a long indwelling time period that required unconventional retrieval techniques.
MATERIALS AND METHODS: We retrospectively reviewed patients who underwent OptEase filter retrieval with long undwelling times requiring additional maneuvers for retrieval. Techniques used included rigid endobronchial forceps dissection and wire-through-loop snare. Each patient underwent postretrieval venogram to evaluate for possible complications. In addition, patients had clinical follow-up 2 weeks after the retrieval procedure.
RESULTS: There were three patients (2 women, 1 man; average age 64 years) who underwent OptEase filter retrieval. The mean indwelling time was 6.4 months. The indwelling filters were successfully retrieved. There were no complications. Postprocedural follow-up showed no clinical pathology.
CONCLUSION: Unconventional techniques aided in the retrieval of OptEase filters with long indwelling times.

Entities:  

Mesh:

Year:  2013        PMID: 23580122     DOI: 10.1007/s00270-013-0619-x

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  3 in total

1.  An analysis of factors associated with increased fluoroscopy time or the need for complex techniques at IVC filter retrieval.

Authors:  Mark Kleedehn; Kelli Moore; Katherine Longo; Kaitlin Woo; Paul Laeseke
Journal:  Eur Radiol       Date:  2018-10-09       Impact factor: 5.315

2.  Bidirectional Pull-Back Technique for Retrieval of Strut-Embedded Cylinder-Shaped Filters in Inferior Vena Cava.

Authors:  Xiaolong Du; Chenglong Li; Aimin Qian; Yeqing Zhang; Wendong Li; Huiying Yu; Xiaoqiang Li; Lingshang Kong
Journal:  Med Sci Monit       Date:  2017-06-09

3.  Pathological Significance of Tissue Trapped in Retrieved OptEase Vena Cava Filter at 18th Day after Installation in Patient with Deep Vein Thrombosis.

Authors:  Sang Min Park; Jung Woo Son; Kyung Soon Hong; Kyung Chan Choi
Journal:  Yonsei Med J       Date:  2017-05       Impact factor: 2.759

  3 in total