Computation of thyroid doses and carcinogenic radiation risks to patients undergoing neck CT examinations.

Walter Huda1, Maria V Spampinato, Sameer V Tipnis, Dennise Magill.   

Abstract

The aim of the study was to investigate how differences in patient anatomy and CT technical factors in neck CT impact on thyroid doses and the corresponding carcinogenic risks. The CTDIvol and dose-length product used in 11 consecutive neck CT studies, as well as data on automatic exposure control (AEC) tube current variation(s) from the image DICOM header, were recorded. For each CT image that included the thyroid, the mass equivalent water cylinder was estimated based on the patient cross-sectional area and average relative attenuation coefficient (Hounsfield unit, HU). Patient thyroid doses were estimated by accounting for radiation intensity at the location of the patient's thyroid, patient size and the scan length. Thyroid doses were used to estimate thyroid cancer risks as a function of patient demographics using risk factors in BEIR VII. The length of the thyroid glands ranged from 21 to 54 mm with an average length of 42 ± 12 mm. Water cylinder diameters corresponding to the central slice through the patient thyroid ranged from 18 to 32 cm with a mean of 25 ± 5 cm. The average CTDIvol (32-cm phantom) used to perform these scans was 26 ± 6 mGy, but the use of an AEC increased the tube current by an average of 44 % at the thyroid mid-point. Thyroid doses ranged from 29 to 80 mGy, with an average of 55 ± 19 mGy. A 20-y-old female receiving the highest thyroid dose of 80 mGy would have a thyroid cancer risk of nearly 0.1 %, but radiation risks decreased very rapidly with increasing patient age. The key factors that affect thyroid doses in neck CT examinations are the radiation intensity at the thyroid location and the size of the patient. The corresponding patient thyroid cancer risk is markedly influenced by patient sex and age.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23579263     DOI: 10.1093/rpd/nct090

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  2 in total

1.  Thyroid doses and risk to paediatric patients undergoing neck CT examinations.

Authors:  Maria Vittoria Spampinato; Sameer Tipnis; Joshua Tavernier; Walter Huda
Journal:  Eur Radiol       Date:  2015-02-02       Impact factor: 5.315

2.  Radiation dose in the thyroid and the thyroid cancer risk attributable to CT scans for pediatric patients in one general hospital of China.

Authors:  Yin-Ping Su; Hao-Wei Niu; Jun-Bo Chen; Ying-Hua Fu; Guo-Bing Xiao; Quan-Fu Sun
Journal:  Int J Environ Res Public Health       Date:  2014-03-07       Impact factor: 3.390

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.