Blood viscometer applying electromagnetically spinning method.

Kazuyoshi Fukunaga1, Masaya Onuki, Yoshinori Ohtsuka, Taichi Hirano, Keiji Sakai, Yasuharu Ohgoe, Ayako Katoh, Toshiyuki Yaguchi, Akio Funakubo, Yasuhiro Fukui.   

Abstract

Viscosity is an important parameter which affects hemodynamics during extracorporeal circulation and long-term cardiac support. In this study, we have aimed to develop a novel viscometer with which we can easily measure blood viscosity by applying the electromagnetically spinning (EMS) method. In the EMS method, we can rotate an aluminum ball 2 mm in diameter indirectly in a test tube with 0.3 ml sample of a liquid by utilizing the moment caused by the Lorentz force as well as separate the test tube from the viscometer body. First, we calibrated the EMS viscometer by means of liquid samples with known viscosities and computational fluid dynamics. Then, when we measured the viscosity of 9.4 mPa s silicone oil in order to evaluate the performance of the EMS viscometer, the mean viscosity was found to be 9.55 ± 0.10 mPa s at available shear rates from 10 to 240 s(-1). Finally, we measured the viscosity of bovine blood. We prepared four blood samples whose hematocrit levels were adjusted to 23, 45, 50, and 70% and a plasma sample without hemocyte components. As a result, the measurements of blood viscosities showed obedience to Casson's equation. We found that the viscosity was approximately constant in Newtonian silicone oil, whereas the viscosity decreased with increasing shear rate in non-Newtonian bovine blood. These results suggest that the EMS viscometer will be useful to measure blood viscosity at the clinical site.

Entities:  

Mesh:

Year:  2013        PMID: 23575974     DOI: 10.1007/s10047-013-0707-3

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  8 in total

1.  Influence of flow properties of blood upon viscosity-hematocrit relationships.

Authors:  R E WELLS; E W MERRILL
Journal:  J Clin Invest       Date:  1962-08       Impact factor: 14.808

2.  An equation for the flow of blood, plasma and serum through glass capillaries.

Authors:  G W BLAIR
Journal:  Nature       Date:  1959-02-28       Impact factor: 49.962

3.  Measurement of whole blood viscosity profiles via an automated viscometer: technical details and clinical relevance.

Authors:  Tamas Alexy; Eszter Pais; Rosalinda B Wenby; William Hogenauer; Kalman Toth; Herbert J Meiselman; Kenneth R Kensey
Journal:  Clin Lab       Date:  2005       Impact factor: 1.138

4.  SIMPLIFIED ROTATING CYLINDER VISCOMETER FOR DNA.

Authors:  B H Zimm; D M Crothers
Journal:  Proc Natl Acad Sci U S A       Date:  1962-06       Impact factor: 11.205

5.  Measurement of rheologic property of blood by a falling-ball blood viscometer.

Authors:  Yoko Eguchi; Takeshi Karino
Journal:  Ann Biomed Eng       Date:  2008-02-08       Impact factor: 3.934

6.  Transitorized falling ball viscometer for rapid testing of biological and technological systems.

Authors:  J L Martin; R Sinusas; R M Jacobs; A L Copley
Journal:  Biorheology       Date:  1974-11       Impact factor: 1.875

7.  Rotational viscometer for powders and granules.

Authors:  C F Harwood; N Pilpel
Journal:  Lab Pract       Date:  1968-11

8.  Fibrinogen, viscosity, and white blood cell count are major risk factors for ischemic heart disease. The Caerphilly and Speedwell collaborative heart disease studies.

Authors:  J W Yarnell; I A Baker; P M Sweetnam; D Bainton; J R O'Brien; P J Whitehead; P C Elwood
Journal:  Circulation       Date:  1991-03       Impact factor: 29.690

  8 in total
  3 in total

1.  Quantitative Correlation between Viscosity of Concentrated MAb Solutions and Particle Size Parameters Obtained from Small-Angle X-ray Scattering.

Authors:  Masakazu Fukuda; Chifumi Moriyama; Tadao Yamazaki; Yoshimi Imaeda; Akiko Koga
Journal:  Pharm Res       Date:  2015-06-16       Impact factor: 4.200

Review 2.  Journal of Artificial Organs 2013: the year in review : Journal of Artificial Organs Editorial Committee.

Authors:  Y Sawa; E Tatsumi; T Tsukiya; K Matsuda; K Fukunaga; A Kishida; T Masuzawa; G Matsumiya; A Myoui; M Nishimura; T Nishimura; T Nishinaka; E Okamoto; S Tokunaga; T Tomo; Y Yagi; T Yamaoka
Journal:  J Artif Organs       Date:  2014-02-26       Impact factor: 1.731

3.  Design and Fabrication of a Microfluidic Viscometer Based on Electrofluidic Circuits.

Authors:  Bo-Bi Tzeng; Yung-Shin Sun
Journal:  Micromachines (Basel)       Date:  2018-07-27       Impact factor: 2.891

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.