| Literature DB >> 23574679 |
Daisuke Jinya1, Tomomi Iwamura, Kiwao Kadokami.
Abstract
A disk-type solid-phase extraction method for simultaneous extraction of a wide range of semi-volatile organic compounds in water samples was developed. The developed method could extract 186 out of 202 model compounds from spiked environmental samples (0.1 μg/L) with 94% average recovery (50-111%). Compared with the previous cartridge-type method, the extraction time was reduced to approximately 1/6. Applying the combination of the developed method and the database system to environmental samples for non-target analyses, 39 micro-pollutants with a wide range of physicochemical properties were determined.Entities:
Year: 2013 PMID: 23574679 DOI: 10.2116/analsci.29.483
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081