| Literature DB >> 23573789 |
Rongfu Huang1, Wei Wang, Yichao Qian, Jessica M Boyd, Yuli Zhao, Xing-Fang Li.
Abstract
We report here the characterization of twelve halobenzoquinones (HBQs) using electrospray ionization (ESI) high resolution quadrupole time-of-flight mass spectrometry. The high resolution negative ESI spectra of the twelve HBQs formed two parent ions, [M + H(+) + 2e(-)], and the radical M(-•). The intensities of these two parent ions are dependent on their chemical structures and on instrumental parameters such as the source temperature and flow rate. The characteristic ions of the HBQs were used to develop an ultra pressure liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method. At the UPLC flow rate (400 μL/min) and under the optimized ESI conditions, eleven HBQs showed the stable and abundant transitions [M + H(+) + 2e(-)] → X(-) (X(-) representing Cl(-), Br(-), or I(-)), while dibromo-dimethyl-benzoquinone (DBDMBQ) showed only the transition of M(-•) → Br(-). The UPLC efficiently separates all HBQs including some HBQ isomers, while the MS/MS offers exquisite limits of detection (LODs) at subng/mL levels for all HBQs except DBDMBQ. Combined with solid phase extraction (SPE), the method LOD is down to ng/L. The results from analysis of authentic samples demonstrated that the SPE-UPLC-MS/MS method is reliable, fast, and sensitive for the identification and quantification of the twelve HBQs in drinking water.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23573789 DOI: 10.1021/ac400160r
Source DB: PubMed Journal: Anal Chem ISSN: 0003-2700 Impact factor: 6.986