Literature DB >> 23572109

Choroidal oximetry with a noninvasive spectrophotometric oximeter.

Jona Valgerdur Kristjansdottir1, Sveinn Hakon Hardarson, Andrew R Harvey, Olof Birna Olafsdottir, Thorunn Scheving Eliasdottir, Einar Stefánsson.   

Abstract

PURPOSE: The purpose of the study was to establish a new technology to measure hemoglobin oxygen saturation in human choroidal vasculature with a noninvasive spectrophotometric oximeter.
METHODS: The fundus camera-based oximeter captures dual-wavelength oximetry images of the fundus and calculates optical density ratio (ODR), which is inversely related to hemoglobin oxygen saturation. Sixteen healthy and lightly pigmented individuals were imaged during normoxia and six during both normoxia and pure oxygen breathing (hyperoxia). ODR was measured for choroidal vessels, vortex veins, and retinal arterioles and venules.
RESULTS: ODR was 0.10 ± 0.10 (mean ± SD) for choroidal vessels, 0.13 ± 0.12 for vortex veins, 0.22 ± 0.04 for retinal arterioles, and 0.50 ± 0.09 for retinal venules. Inhalation of pure oxygen lowered ODR levels in all vessel types; the decrease was 0.035 ± 0.028 in choroidal vessels (P = 0.029, paired t-test), 0.022 ± 0.017 in the retinal arterioles (P = 0.022, paired t-test), and 0.246 ± 0.067 in retinal venules (P = 0.0003, paired t-test).
CONCLUSIONS: The ODR can be measured noninvasively in the choroidal vessels of lightly pigmented individuals and is significantly lower in choroidal vessels than in retinal arterioles. This may suggest higher oxygen saturation but is also compatible with the reduced contrast of choroidal vessels at both wavelengths that is expected from scattering of light within the choroid. The decrease of ODR during hyperoxia was significant for all vessel types, which confirms that the oximeter is sensitive to changes in oxygen saturation in both choroidal and retinal vessels.

Entities:  

Keywords:  choroid; hyperoxia; optical density; oxygen; retina

Mesh:

Substances:

Year:  2013        PMID: 23572109     DOI: 10.1167/iovs.12-10507

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  6 in total

Review 1.  Paradigm Shifts in Ophthalmic Diagnostics.

Authors:  J Sebag; Alfredo A Sadun; Eric A Pierce
Journal:  Trans Am Ophthalmol Soc       Date:  2016-08

2.  Monte-Carlo simulation and tissue-phantom model for validation of ocular oximetry.

Authors:  Cléophace Akitegetse; Patricia Landry; Jonathan Robidoux; Nicolas Lapointe; Danny Brouard; Dominic Sauvageau
Journal:  Biomed Opt Express       Date:  2022-04-21       Impact factor: 3.562

3.  Hyperspectral Image Mapping Spectrometry for Retinal Oximetry Measurements in Four Diseased Eyes.

Authors:  Jason G Dwight; Christina Y Weng; Robert E Coffee; Michal E Pawlowski; Tomasz S Tkaczyk
Journal:  Int Ophthalmol Clin       Date:  2016

4.  Molecular probes for imaging of hypoxia in the retina.

Authors:  Stephanie M Evans; Kwangho Kim; Chauca E Moore; Md Imam Uddin; Megan E Capozzi; Jason R Craft; Gary A Sulikowski; Ashwath Jayagopal
Journal:  Bioconjug Chem       Date:  2014-10-09       Impact factor: 4.774

5.  Retinal and choroidal oxygen saturation of the optic nerve head in open-angle glaucoma subjects by multispectral imaging.

Authors:  Gai-Yun Li; Samer Abdo Al-Wesabi; Hong Zhang
Journal:  Medicine (Baltimore)       Date:  2016-12       Impact factor: 1.889

Review 6.  Advances in Retinal Oximetry.

Authors:  Anupam K Garg; Darren Knight; Leonardo Lando; Daniel L Chao
Journal:  Transl Vis Sci Technol       Date:  2021-02-05       Impact factor: 3.283

  6 in total