| Literature DB >> 23571750 |
Abstract
Readjustable suburethral sling procedures have become established as a standard method for therapy of postoperative urinary stress incontinence in men. Due to the silicone construction revision after implantation of Argus ™ slings can be carried out without problems even after a long indwell time. In the case presented correction of sling-related hydronephrosis due to incorrect positioning of the Argus ™ sling is demonstrated. Surgical correction was possible without explantation or exchanging the system and ensuring a good functional outcome.Entities:
Mesh:
Year: 2013 PMID: 23571750 DOI: 10.1007/s00120-013-3168-7