Literature DB >> 23567628

High-κ GdTixOy sensing membrane-based electrolyte-insulator-semiconductor with magnetic nanoparticles as enzyme carriers for protein contamination-free glucose biosensing.

Min-Hsien Wu1, Hung-Wei Yang, Mu-Yi Hua, Yen-Bo Peng, Tung-Ming Pan.   

Abstract

This paper reports an electrolyte-insulator-semiconductor (EIS) device featuring a novel high-κ GdTixOy sensing membrane for high-performance pH sensing and glucose biosensing. The effect of the annealing temperature (700, 800, or 900°C) on the sensing properties of the GdTixOy membranes was investigated. The GdTixOy EIS device annealed at 900°C exhibited the greatest pH sensing performance, including the highest sensitivity (62.12mV/pH), the smallest hysteresis voltage (5mV), and the lowest drift rate (0.4mV/h), presumably because of its well-crystallized GdTixOy structure. To overcome the problems typically encountered during the practical application of biosensors (e.g., protein adsorption; preservation of enzymatic activity), we employed Fe3O4-based magnetic nanoparticles (MNPs) as enzyme carriers. The adsorption of serum protein on the unmodified sensing membrane led to poor EIS-based pH sensing (r(2)=0.71); the performance was greatly improved, however, after attaching the MNPs to the sensing membrane, thereby blocking protein adsorption significantly (by 98%) and allowing excellent pH sensing (r(2)=0.99). Moreover, we prepared a hybrid configuration of the proposed GdTixOy membrane-EIS, with magnetically attached glucose oxidase-immobilized MNPs, for glucose biosensing. The use of MNPs as enzyme carriers effectively preserved the enzymatic activity of glucose oxidase, with 45.3% of the original enzymatic activity retained after 120h of storage at 4°C (compared with complete loss of the free enzyme's activity under the same storage conditions). In addition, the proposed biosensor exhibited superior detection sensitivity of 11.03mV/mM relative to that (8.17mV/mM) obtained using the conventional enzyme immobilization method. Finally, we established the accuracy of the proposed method for blood glucose measurement; gratifyingly, blood glucose detection was comparable with the high-sensitivity glucose quantification obtained using a commercial glucose assay kit.
Copyright © 2013 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23567628     DOI: 10.1016/j.bios.2013.03.013

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  1 in total

Review 1.  Capacitive Field-Effect EIS Chemical Sensors and Biosensors: A Status Report.

Authors:  Arshak Poghossian; Michael J Schöning
Journal:  Sensors (Basel)       Date:  2020-10-02       Impact factor: 3.576

  1 in total