Literature DB >> 23565714

Calibration and validation of nine objective vision screeners with contact lens-induced anisometropia.

Robert W Arnold1, Bethanne Davis, Laura E Arnold, Kayla S Rowe, Jodi M Davis.   

Abstract

PURPOSE: New objective photoscreeners are emerging with encouraging national guidelines and pediatric providers can be reimbursed for photoscreening.
METHODS: Nine objective screeners were applied to two emmetropic subjects with or without known power contact lenses to induce spherical and cylindrical anisometropia close to American Association for Pediatric Ophthalmology and Strabismus (AAPOS) thresholds.
RESULTS: The screeners produced near linear pupil crescents and estimated refractions for induced anisometropia. Most performed well with AAPOS 2003 validation; however, the iPhone (Apple, Cupertino, CA) was too sensitive, whereas the remote autorefractors using Vision in Preschoolers Study criteria were less sensitive.
CONCLUSIONS: The new, interpreted objective screeners appear valid for detection of anismetropia. Copyright 2013, SLACK Incorporated.

Entities:  

Mesh:

Year:  2013        PMID: 23565714     DOI: 10.3928/01913913-20130402-02

Source DB:  PubMed          Journal:  J Pediatr Ophthalmol Strabismus        ISSN: 0191-3913            Impact factor:   1.402


  6 in total

1.  The effectiveness of the Spot Vision Screener in detecting amblyopia risk factors.

Authors:  Mae Millicent W Peterseim; Carrie E Papa; M Edward Wilson; Jennifer D Davidson; Maria Shtessel; Mavesh Husain; Edward W Cheeseman; Bethany J Wolf; Rupal Trivedi
Journal:  J AAPOS       Date:  2014-12       Impact factor: 1.220

2.  Photoscreeners in the pediatric eye office: compared testability and refractions on high-risk children.

Authors:  Mae Millicent W Peterseim; Carrie E Papa; M Edward Wilson; Edward W Cheeseman; Bethany J Wolf; Jennifer D Davidson; Rupal H Trivedi
Journal:  Am J Ophthalmol       Date:  2014-08-01       Impact factor: 5.258

3.  Grading Sphero-Cylinder Spectacle Similarity.

Authors:  Robert W Arnold; Joshua S Beveridge; Samuel J Martin; Nathanael R Beveridge; Elise J Metzger; Kyle A Smith
Journal:  Clin Optom (Auckl)       Date:  2021-01-20

Review 4.  Scope and costs of autorefraction and photoscreening for childhood amblyopia-a systematic narrative review in relation to the EUSCREEN project data.

Authors:  Anna M Horwood; Helen J Griffiths; Jill Carlton; Paolo Mazzone; Arinder Channa; Mandy Nordmann; Huibert J Simonsz
Journal:  Eye (Lond)       Date:  2020-11-30       Impact factor: 3.775

5.  Comparative Validation of PlusoptiX and AI-Optic Photoscreeners in Children with High Amblyopia Risk Factor Prevalence.

Authors:  Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2022-08-16

6.  Performance of Two Photoscreeners Enhanced by Protective Cases.

Authors:  Samuel Joseph Martin; Hser Eh Htoo; Nay Hser; Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2020-05-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.