Quantitative characterization of tomotherapy MVCT dosimetry.

Mingli Chen1, Edward Chao, Weiguo Lu.   

Abstract

Megavoltage computed tomography (MVCT) is used as image guidance for patient setup in almost every tomotherapy treatment. Frequent use of ionizing radiation for image guidance has raised concern of imaging dose. The purpose of this work is to quantify and characterize tomotherapy MVCT dosimetry. Our dose calculation was based on a commissioned dose engine, and the calculation result was compared with film measurement. We studied dose profiles, center dose, maximal dose, surface dose, and mean dose on homogeneous cylindrical water phantoms of various diameters for various scanning parameters, including 3 different jaw openings (of nominal value J4, J1, and J0.1) and couch speeds (fine, normal, and coarse). The comparison between calculation and film measurement showed good agreement. In particular, the thread pattern on the film of the helical delivery matched very well with calculation. For the J1 jaw and coarse imaging mode, the maximum difference between calculation and measurement was about 6% of the center dose. Calculation on various sizes of synthesized phantoms showed that the center dose decreases almost linearly as the phantom diameter increases, and that the fine mode (couch speed of 4mm/rotation) received twice the dose of the normal mode (couch speed of 8mm/rotation) and 3 times that of the coarse mode (couch speed of 12mm/rotation) as expected. The maximal dose ranged from 100% to ∼200% of the center dose, with increasing ratios for larger phantoms, smaller jaws, and faster couch speed. For all jaw settings and couch speeds, the mean dose and average surface dose vary from 95% to 125% of the center dose with increasing ratios for larger phantoms. We present a quantitative dosimetric characterization of the tomotherapy MVCT in terms of scanning parameters, phantom size, center dose, maximal dose, surface dose, and mean dose. The results can provide an overall picture of dose distribution and a reference data set that enables estimation of CT dose index for the tomotherapy MVCT.
Copyright © 2013 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  MVCT dose quantification; Tomotherapy

Mesh:

Year:  2013        PMID: 23558147     DOI: 10.1016/j.meddos.2013.02.009

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  2 in total

1.  Evaluation of MVCT imaging dose levels during helical IGRT: comparison between ion chamber, TLD, and EBT3 films.

Authors:  Jean-Pierre Mege; Sun Wenzhao; Attila Veres; Guillaume Auzac; Ibrahima Diallo; Dimitri Lefkopoulos
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

2.  Effects of megavoltage computed tomographic scan methodology on setup verification and adaptive dose calculation in helical TomoTherapy.

Authors:  Jian Zhu; Tong Bai; Jiabing Gu; Ziwen Sun; Yumei Wei; Baosheng Li; Yong Yin
Journal:  Radiat Oncol       Date:  2018-04-27       Impact factor: 3.481

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.