Dose response of selected ion chambers in applied homogeneous transverse and longitudinal magnetic fields.

M Reynolds1, B G Fallone, S Rathee.   

Abstract

PURPOSE: The magnetic fields of an integrated MR-Linac system will alter the paths of electrons that produce ions in the ionization chambers. The dose response of selected ion chambers is evaluated in the presence of varying transverse and longitudinal magnetic fields. The investigation is useful in calibration of therapeutic x-ray beams associated with MR-Linac systems.
METHODS: The Monte Carlo code PENELOPE was used to model the irradiation of NE2571, and PR06C ionization chambers in the presence of a transverse and longitudinal (with respect to the photon beam) magnetic fields of varying magnitude. The long axis of each chamber was simulated both parallel and perpendicular to the incident photon beam for each magnetic field case. The dose deposited in each chamber for each case was compared to the case with zero magnetic field by means of a ratio. The PR06C chamber's response was measured in the presence of a transverse magnetic field with field strengths ranging from 0.0 to 0.2 T to compare to simulated results.
RESULTS: The simulations and measured data show that in the presence of a transverse magnetic field there is a considerable dose response (maximum of 11% near 1.0 T in the ion chambers investigated, which depends on the magnitude of magnetic field, and relative orientation of the magnetic field, radiation beam, and ion chamber. Measurements made with the PR06C chamber verify these results in the region of measurement. In contrast, a longitudinal magnetic field produces only a slight increase in dose response (2% at 1.5 T) that rises slowly with increasing magnetic field and is seemingly independent of chamber orientation. Response trends were similar for the two ion chambers and relative orientations considered, but slight variations are present from chamber to chamber.
CONCLUSIONS: Care must be taken when making ion chamber measurements in a transverse magnetic field. Ion chamber responses vary not only with transverse field strength, but with chamber orientation and type, and can be considerable. Longitudinal magnetic fields influence ion chamber responses relatively little (2% at 1.5 T), and only at field strengths in excess of 1.0 T.

Entities:  

Mesh:

Year:  2013        PMID: 23556912     DOI: 10.1118/1.4794496

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

Review 1.  A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy.

Authors:  S O'Keeffe; D McCarthy; P Woulfe; M W D Grattan; A R Hounsell; D Sporea; L Mihai; I Vata; G Leen; E Lewis
Journal:  Br J Radiol       Date:  2015-03-11       Impact factor: 3.039

2.  Relative dosimetry with an MR-linac: Response of ion chambers, diamond, and diode detectors for off-axis, depth dose, and output factor measurements.

Authors:  Daniel J O'Brien; James Dolan; Stefan Pencea; Nicholas Schupp; Gabriel O Sawakuchi
Journal:  Med Phys       Date:  2017-12-21       Impact factor: 4.071

3.  Effect of Magnetic Field Strength on Plastic Scintillation Detector Response.

Authors:  F Therriault-Proulx; Z Wen; G Ibbott; S Beddar
Journal:  Radiat Meas       Date:  2018-06-09       Impact factor: 1.898

Review 4.  Optical fibre sensors: their role in in vivo dosimetry for prostate cancer radiotherapy.

Authors:  P Woulfe; F J Sullivan; S O'Keeffe
Journal:  Cancer Nanotechnol       Date:  2016-10-18

5.  An analysis of the ArcCHECK-MR diode array's performance for ViewRay quality assurance.

Authors:  Steven T Ellefson; Wesley S Culberson; Bryan P Bednarz; Larry A DeWerd; John E Bayouth
Journal:  J Appl Clin Med Phys       Date:  2017-06-06       Impact factor: 2.102

6.  Technical Note: Consistency of PTW30013 and FC65-G ion chamber magnetic field correction factors.

Authors:  S J Woodings; B van Asselen; T L van Soest; L A de Prez; J J W Lagendijk; B W Raaymakers; J W H Wolthaus
Journal:  Med Phys       Date:  2019-06-17       Impact factor: 4.071

7.  Impact of magnetic fields on dose measurement with small ion chambers illustrated in high-resolution response maps.

Authors:  Joerg Lehmann; Toby Beveridge; Chris Oliver; Tracy E Bailey; Jessica E Lye; Jayde Livingstone; Andrew W Stevenson; Duncan J Butler
Journal:  Med Phys       Date:  2019-06-11       Impact factor: 4.071

Review 8.  Medical physics challenges in clinical MR-guided radiotherapy.

Authors:  Christopher Kurz; Giulia Buizza; Guillaume Landry; Florian Kamp; Moritz Rabe; Chiara Paganelli; Guido Baroni; Michael Reiner; Paul J Keall; Cornelis A T van den Berg; Marco Riboldi
Journal:  Radiat Oncol       Date:  2020-05-05       Impact factor: 3.481

9.  Characterization of EBT3 radiochromic films for dosimetry of proton beams in the presence of magnetic fields.

Authors:  Fatima Padilla-Cabal; Peter Kuess; Dietmar Georg; Hugo Palmans; Lukas Fetty; Hermann Fuchs
Journal:  Med Phys       Date:  2019-05-31       Impact factor: 4.071

10.  Characterizing magnetically focused contamination electrons by off-axis irradiation on an inline MRI-Linac.

Authors:  Elizabeth Patterson; Bradley M Oborn; Dean Cutajar; Urszula Jelen; Gary Liney; Anatoly B Rosenfeld; Peter E Metcalfe
Journal:  J Appl Clin Med Phys       Date:  2022-03-25       Impact factor: 2.243