Finite element analysis of an implant-assisted removable partial denture.

Reza Shahmiri1, John M Aarts2, Vincent Bennani1, Momen A Atieh2, Michael V Swain2.   

Abstract

PURPOSE: This study analyzes the effects of loading a Kennedy class I implant-assisted removable partial denture (IARPD) using finite element analysis (FEA). Standard RPDs are not originally designed to accommodate a posterior implant load point. The null hypothesis is that the introduction of posteriorly placed implants into an RPD has no effect on the load distribution.
MATERIALS AND METHODS: A Faro Arm scan was used to extract the geometrical data of a human partially edentulous mandible. A standard plus regular neck (4.8 × 12 mm) Straumann® implant and titanium matrix, tooth roots, and periodontal ligaments were modeled using a combination of reverse engineering in Rapidform XOR2 and solid modeling in Solidworks 2008 FEA program. The model incorporated an RPD and was loaded with a bilateral force of 120 N. ANSYS Workbench 11.0 was used to analyze deformation in the IARPD and elastic strain in the metal framework.
RESULTS: FEA identified that the metal framework developed high strain patterns on the major and minor connectors, and the acrylic was subjected to deformation, which could lead to acrylic fractures. The ideal position of the neutral axis was calculated to be 0.75 mm above the ridge.
CONCLUSION: A potentially destructive mismatch of strain distribution was identified between the acrylic and metal framework, which could be a factor in the failure of the acrylic. The metal framework showed high strain patterns on the major and minor connectors around the teeth, while the implant components transferred the load directly to the acrylic.
© 2013 by the American College of Prosthodontists.

Entities:  

Keywords:  Implant-assisted removable partial denture; dental implant; finite element analysis; removable partial denture

Mesh:

Substances:

Year:  2013        PMID: 23551664     DOI: 10.1111/jopr.12031

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  4 in total

1.  Three-dimensional isotropic magnetic resonance imaging can provide a reliable estimate of the native anterior cruciate ligament insertion site anatomy.

Authors:  Daisuke Araki; Eric Thorhauer; Scott Tashman
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-06-13       Impact factor: 4.342

2.  A three-dimensional finite element analysis of mechanical function for 4 removable partial denture designs with 3 framework materials: CoCr, Ti-6Al-4V alloy and PEEK.

Authors:  Xin Chen; Bochun Mao; Zhuoli Zhu; Jiayi Yu; Yuqing Lu; Qianqian Zhang; Li Yue; Haiyang Yu
Journal:  Sci Rep       Date:  2019-09-27       Impact factor: 4.379

3.  Stress Distribution Pattern in Mini Dental Implant-Assisted RPD with Different Clasp Designs: 3D Finite Element Analysis.

Authors:  Chaiy Rungsiyakull; Pimduen Rungsiyakull; Kullapop Suttiat; Nut Duangrattanaprathip
Journal:  Int J Dent       Date:  2022-03-11

4.  Effect of the Location of Dental Mini-Implants on Strain Distribution under Mandibular Kennedy Class I Implant-Retained Removable Partial Dentures.

Authors:  Pimduen Rungsiyakull; Kallaya Kujarearntaworn; Pathawee Khongkhunthian; Michael Swain; Chaiy Rungsiyakull
Journal:  Int J Dent       Date:  2021-05-04