Literature DB >> 23551597

Cellular origin and diagnostic significance of high-fluorescent cells in cerebrospinal fluid detected by the XE-5000 hematology analyzer.

M Zimmermann1, C Otto, J-B Gonzalez, S Prokop, K Ruprecht.   

Abstract

INTRODUCTION: The Sysmex XE-5000 is a blood and body fluid analyzer able to differentiate cells into polymorphonuclear, mononuclear, and high-fluorescent cells (HFC). The identity of HFC in cerebrospinal fluid (CSF) has been uncertain; however, compatible with their high nucleic acid content, HFC could represent intrathecal tumor cells. Here, we studied the cellular origin and the diagnostic significance of HFC in CSF.
METHODS: Results of CSF examinations with the XE-5000 were analyzed in 65 CSF samples with and 126 CSF samples without tumor cells, as defined by manual microscopy of CSF cytospin preparations.
RESULTS: The XE-5000 detected HFC in 51 of 65 tumor cell-positive and in 33 of 126 tumor cell-negative CSF samples (sensitivity: 78.5%, specificity: 73.8%, positive likelihood ratio: 3.0, negative likelihood ratio: 0.29). The percentages of HFC and tumor cells in CSF samples correlated (r² = 0.41, P < 0.0001). Tumor cells escaped detection by the XE-5000 especially in CSF samples with a low percentage of tumor cells.
CONCLUSION: While this study identifies tumor cells as the predominant correlate of HFC in CSF, it suggests that measuring HFC is not an appropriate diagnostic test for intrathecal tumor cells. However, if HFC are incidentally detected in CSF, further evaluation by CSF microscopy seems mandatory.
© 2013 John Wiley & Sons Ltd.

Entities:  

Keywords:  CSF; cytospin preparations; high-fluorescent cells; tumor cells

Mesh:

Year:  2013        PMID: 23551597     DOI: 10.1111/ijlh.12090

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Screening for malignant tumor cells in serous effusions with an automatic hematology analyzer using a novel diagnostic algorithm.

Authors:  Dehua Sun; Xinru Mao; Taixue An; Xiaojing He; Kai Qiu; Yuhong Luo; Zheyuan Qin; Yongjian He; Tie Xiong; Houmei Feng; Jin Li; Lei Zheng
Journal:  Ann Transl Med       Date:  2022-03

2.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

  2 in total