BACKGROUND: To facilitate the correct use of epinephrine autoinjectors (EAIs) by patients and caregivers, a novel EAI (Auvi-Q) was designed to help minimize use-related hazards.
OBJECTIVE: To support validation of Auvi-Q final design and assess whether the instructions for use in the patient information leaflet (PIL) are effective in training participants on proper use of Auvi-Q.
METHODS: Healthy participants, 20 adult and 20 pediatric, were assessed for their ability to complete a simulated injection by following the Auvi-Q instructions for use. Participants relied only on the contents of the PIL and other labeling features (device labeling and its instructions for use, electronic voice instructions and visual prompts).
RESULTS: The mean ± SD age of the adult and pediatric participants was 39.4 ± 11.6 and 10.9 ± 2.3 years, respectively. In total, 80% of adult and 35% of pediatric participants had prior experience with EAIs. All adults and 95% of pediatric participants completed a simulated injection on the first attempt; 1 pediatric participant required parental training and a second attempt. Three adult and 4 pediatric participants exhibited a noncritical issue while successfully completing the simulated injection. Most participants agreed that the injection steps were easy to follow and the PIL facilitated understanding on using Auvi-Q safely and effectively.
CONCLUSION: The PIL and other labeling features were effective in communicating instructions for successful use of Auvi-Q. This study provided validation support for the final design and anticipated instructions for use of Auvi-Q.
Copyright © 2013 American College of Allergy, Asthma & Immunology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23548530     DOI: 10.1016/j.anai.2012.12.008

Source DB:  PubMed          Journal:  Ann Allergy Asthma Immunol        ISSN: 1081-1206            Impact factor:   6.347


  1 in total

1.  Usability of a new disposable autoinjector platform device: results of a formative study conducted with a broad user population.

Authors:  Jakob Lange; Philipp Richard; Nick Bradley
Journal:  Med Devices (Auckl)       Date:  2015-06-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.