The effect of rotational fixation error of the tibial cutting guide and the distance between the guide and the bone on the tibial osteotomy in total knee arthroplasty.

Tadashi Tsukeoka1, Yoshikazu Tsuneizumi, Tae Hyun Lee.   

Abstract

Computed tomography based computer simulation studies were made on fifty consecutive patients with osteoarthritis scheduled for TKA. Proximal tibial cutting was simulated with the cutting guide placed at different rotational errors (-10°, 0°, 10° and 20° of internal rotation) and different distances (4 cm and 8 cm) between the alignment rod and the bone. Only ten degrees of rotation error with 8 cm of distance resulted in over 3° of varus or valgus tibial cut. Our study demonstrated that rotational mismatch between the proximal and distal part of the tibia should be avoided and the distance between the alignment rod and the bone should be as short as possible to achieve proper tibial alignment.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  extramedullary cutting guide; mechanical axis; proximal tibial osteotomy; total knee arthroplasty

Mesh:

Year:  2013        PMID: 23528550     DOI: 10.1016/j.arth.2012.12.008

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Accuracy of the second metatarsal as a landmark for the extramedullary tibial cutting guide in total knee arthroplasty.

Authors:  Tadashi Tsukeoka; Yoshikazu Tsuneizumi; Tae Hyun Lee
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-08-27       Impact factor: 4.342

2.  Rotation of intramedullary alignment rods affects distal femoral cutting plane in total knee arthroplasty.

Authors:  Günther Maderbacher; Jan Matussek; Armin Keshmiri; Felix Greimel; Clemens Baier; Joachim Grifka; Hermann Maderbacher
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2018-02-17       Impact factor: 4.342

  2 in total