Retentive strength of two-piece CAD/CAM zirconia implant abutments.

Peter Gehrke1, Jochen Alius, Carsten Fischer, Kurt J Erdelt, Florian Beuer.   

Abstract

PURPOSE: The purpose of this study is to evaluate the retention of two-piece computer-aided design (CAD)/computer aided manufacturing (CAM) zirconia abutments after artificial aging under simulated oral conditions using three different types of resin-based luting agents.
MATERIAL AND METHODS: Twenty-one CAD/CAM-generated zirconia copings (CERCON Compartis, Degudent, Hanau, Germany) were bonded to a prefabricated secondary titanium implant insert (XiVE Ti-Base, Dentsply Friadent, Mannheim, Germany), using three different types of resin-based luting agents: group A: Panavia 21 (Kuraray Co, Kurashiki, Japan); group B: Multilink Implant (Ivoclar Vivadent, Schaan, Liechtenstein); and group C: SmartCem2 (Dentsply DeTrey, Konstanz, Germany). The bonding surfaces of the titanium inserts and the zirconia ceramic copings were air-abraded and cleaned in alcohol. All specimens were stored in distilled water for 60 days and subsequently thermal-cycled 15,000 times (5-55 °C). The dislodging force of the copings along the long axis of the implant/abutment complex was recorded using a universal testing machine with 2 mm/min crosshead speed. Data were analyzed descriptively and by performing the Kruskal-Wallis test.
RESULTS: The mean retention values were 924.93 ± 363.31 N for Panavia 21, 878.05 ± 208.33 N for Multilink Implant, and 650.77 ± 174.92 N for SmartCem2. The Kruskal-Wallis test indicated no significant difference between the retention values of the tested luting agents (p = 0.1314). The failure modes of all tested two-piece abutments were completely adhesive, leaving the detached zirconia coping and titanium insert undamaged.
CONCLUSION: The use of resin-based luting agents in combination with air abrasion of titanium inserts and zirconia copings led to a stable retention of two-piece CAD/CAM abutments. The bonding stability of the investigated luting agents exceeded the general limits of fracture resistance of two-piece zirconia abutments. A notable difference between the mean retention values of the tested bond materials was shown. However, the statistical analysis revealed that this difference was not significant.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  bonding; computer assisted; titanium; two-piece CAD/CAM abutments; zirconia

Mesh:

Substances:

Year:  2013        PMID: 23527950     DOI: 10.1111/cid.12060

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  8 in total

1.  Application of a standard autoclaving protocol does not harm structural integrity of two-piece zirconia abutments under detachment force testing.

Authors:  Dieter Pils; Ralf J Baeppler; Rüdiger Junker; Andrej M Kielbassa; Frank P Nothdurft
Journal:  Clin Oral Investig       Date:  2019-05-10       Impact factor: 3.573

2.  Mechanical stability of angulated zirconia abutments supporting maxillary anterior single crowns on narrow-diameter implants.

Authors:  Ruiyang Ma; Ping Yu; Yuqiang Zhang; Chenyang Xie; Xin Tan; Jikui Sun; Haiyang Yu
Journal:  Clin Oral Investig       Date:  2022-09-26       Impact factor: 3.606

3.  Microscopical and chemical surface characterization of CAD/CAM zircona abutments after different cleaning procedures. A qualitative analysis.

Authors:  Peter Gehrke; Astrid Tabellion; Carsten Fischer
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

4.  The Use of a Hybrid Pillar and Its Importance for Aesthetic Rehabilitation and Tissue Stability: A Clinical Report.

Authors:  Guilherme da Gama Ramos; Danilo Lazzari Ciotti; Samuel Rehder Wimmers Ferreira; Maide Rehder Wimmers Ferreira Margarido; Raquel Adriano Dantas; Marina Nottingham Guerreiro
Journal:  Case Rep Dent       Date:  2018-07-02

5.  Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Authors:  Mitra Zirak; Mahroo Vojdani; Amir Ali Reza Khaledi; Mitra Farzin
Journal:  J Dent (Shiraz)       Date:  2019-06

6.  Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.

Authors:  Yong-Seok Jang; Sang-Hoon Oh; Won-Suck Oh; Min-Ho Lee; Jung-Jin Lee; Tae-Sung Bae
Journal:  Materials (Basel)       Date:  2019-08-30       Impact factor: 3.623

7.  Titanium Base Abutments in Implant Prosthodontics: A Literature Review.

Authors:  Ahmad M Al-Thobity
Journal:  Eur J Dent       Date:  2021-11-18

8.  Load-Bearing Capacity of Zirconia Crowns Screwed to Multi-Unit Abutments with and without a Titanium Base: An In Vitro Pilot Study.

Authors:  Hadas Heller; Adi Arieli; Ilan Beitlitum; Raphael Pilo; Shifra Levartovsky
Journal:  Materials (Basel)       Date:  2019-09-20       Impact factor: 3.623

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.