Can current analytical quality performance of UK clinical laboratories support evidence-based guidelines for diabetes and ischaemic heart disease?--A pilot study and a proposal.

Nuthar Jassam1, John Yundt-Pacheco, Rob Jansen, Annette Thomas, Julian H Barth.   

Abstract

BACKGROUND: The implementation of national and international guidelines is beginning to standardise clinical practice. However, since many guidelines have decision limits based on laboratory tests, there is an urgent need to ensure that different laboratories obtain the same analytical result on any sample. A scientifically-based quality control process will be a pre-requisite to provide this level of analytical performance which will support evidence-based guidelines and movement of patients across boundaries while maintaining standardised outcomes. We discuss the finding of a pilot study performed to assess UK clinical laboratories readiness to work to a higher grade quality specifications such as biological variation-based quality specifications.
METHODS: Internal quality control (IQC) data for HbA1c, glucose, creatinine, cholesterol and high density lipoprotein (HDL)-cholesterol were collected from UK laboratories participating in the Bio-Rad Unity QC programme. The median of the coefficient of variation (CV%) of the participating laboratories was evaluated against the CV% based on biological variation.
RESULTS: Except creatinine, the other four analytes had a variable degree of compliance with the biological variation-based quality specifications. More than 75% of the laboratories met the biological variation-based quality specifications for glucose, cholesterol and HDL-cholesterol. Slightly over 50% of the laboratories met the analytical goal for HBA1c. Only one analyte (cholesterol) had a performance achieving the higher quality specifications consistent with 5σ.
CONCLUSIONS: Our data from IQC do not consistently demonstrate that the results from clinical laboratories meet evidence-based quality specifications. Therefore, we propose that a graded scale of quality specifications may be needed at this stage.

Entities:  

Mesh:

Year:  2013        PMID: 23525878     DOI: 10.1515/cclm-2012-0840

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  2 in total

Review 1.  Harmonization: the sample, the measurement, and the report.

Authors:  W Greg Miller; Jillian R Tate; Julian H Barth; Graham R D Jones
Journal:  Ann Lab Med       Date:  2014-04-08       Impact factor: 3.464

2.  Creatinine Assay Attainment of Analytical Performance Goals Following Implementation of IDMS Standardization: Further Improvements Required.

Authors:  Elizabeth Sunmin Lee; Christine P Collier; Christine A White
Journal:  Can J Kidney Health Dis       Date:  2017-02-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.