Dosimetric impact of intrafraction motion during RapidArc stereotactic vertebral radiation therapy using flattened and flattening filter-free beams.

Chin Loon Ong1, Max Dahele, Johan P Cuijpers, Suresh Senan, Ben J Slotman, Wilko F A R Verbakel.   

Abstract

PURPOSE: To study the dosimetric impact of relatively short-duration intrafraction shifts during a single fraction of RapidArc delivery for vertebral stereotactic body radiation therapy (SBRT) using flattened (FF) and flattening filter-free (FFF) beams. METHODS AND MATERIALS: The RapidArc plans, each with 2 to 3 arcs, were generated for 9 patients using 6-MV FF and 10-MV FFF beams with maximum dose rates of 1000 and 2400 MU/min, respectively. A total of 1272 plans were created to estimate the dosimetric consequences in target and spinal cord volumes caused by intrafraction shifts during one of the arcs. Shifts of 1, 2, and 3 mm for periods of 5, 10, and 30 seconds, and 5 mm for 5 and 10 seconds, were modelled during a part of the arc associated with high doses and steep dose gradients.
RESULTS: For FFF plans, shifts of 2 mm over 10 seconds and 30 seconds could increase spinal cord Dmax by up to 6.5% and 13%, respectively. Dosimetric deviations in FFF plans were approximately 2-fold greater than in FF plans. Reduction in target coverage was <1% for 83% and 96% of the FFF and FF plans, respectively.
CONCLUSION: Even short-duration intrafraction shifts can cause significant dosimetric deviations during vertebral SBRT delivery, especially when using very high dose rate FFF beams and when the shift occurs in that part of the arc delivering high doses and steep gradients. The impact is greatest on the spinal cord and its planning-at-risk volume. Accurate and stable patient positioning is therefore required for vertebral SBRT.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23523183     DOI: 10.1016/j.ijrobp.2012.12.028

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  6 in total

1.  Effectiveness of robust optimization in volumetric modulated arc therapy using 6 and 10 MV flattening filter-free beam therapy planning for lung stereotactic body radiation therapy with a breath-hold technique.

Authors:  Hideharu Miura; Shuichi Ozawa; Yoshiko Doi; Minoru Nakao; Katsumaro Kubo; Masahiko Kenjo; Yasushi Nagata
Journal:  J Radiat Res       Date:  2020-07-06       Impact factor: 2.724

2.  Investigating the dosimetric effects of grid size on dose calculation accuracy using volumetric modulated arc therapy in spine stereotactic radiosurgery.

Authors:  Chin Snyder Karen; Manju Liu; Bo Zhao; Yimei Huang; Wen Ning; Indrin J Chetty; M Salim Siddiqui
Journal:  J Radiosurg SBRT       Date:  2017

3.  Safety of high-dose-rate stereotactic body radiotherapy.

Authors:  Sonja Stieb; Stephanie Lang; Claudia Linsenmeier; Shaun Graydon; Oliver Riesterer
Journal:  Radiat Oncol       Date:  2015-01-23       Impact factor: 3.481

4.  On the robustness of VMAT-SABR treatment plans against isocentre positioning uncertainties.

Authors:  Joep Stroom; Sandra Vieira; Dalila Mateus; Carlo Greco; Antonella Fogliata; Giorgia Nicolini; Alessandro Clivio; Eugenio Vanetti; Luca Cozzi
Journal:  Radiat Oncol       Date:  2014-09-05       Impact factor: 3.481

5.  Dosimetric differences in flattened and flattening filter-free beam treatment plans.

Authors:  Yue Yan; Poonam Yadav; Michael Bassetti; Kaifang Du; Daniel Saenz; Paul Harari; Bhudatt R Paliwal
Journal:  J Med Phys       Date:  2016 Apr-Jun

6.  Evaluation of a proprietary software application for motion monitoring during stereotactic paraspinal treatment.

Authors:  Qiyong Fan; Hai Pham; Pengpeng Zhang; Xiang Li; Tianfang Li
Journal:  J Appl Clin Med Phys       Date:  2022-03-26       Impact factor: 2.243

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.