| Literature DB >> 23522723 |
Michael R Backhouse1, Elizabeth M A Hensor, Derrick White, Anne-Maree Keenan, Philip S Helliwell, Anthony C Redmond.
Abstract
BACKGROUND: Physical activity is frequently reported in rheumatology but it is difficult to measure objectively outside the gait laboratory. A new generation of activity monitors offers this potential but it has not yet been evaluated in patients with rheumatoid arthritis. This study aimed to evaluate three types of activity monitors in patients with rheumatoid arthritis.Entities:
Mesh:
Year: 2013 PMID: 23522723 PMCID: PMC3677088 DOI: 10.1016/j.clinbiomech.2013.02.009
Source DB: PubMed Journal: Clin Biomech (Bristol, Avon) ISSN: 0268-0033 Impact factor: 2.063
Fig. 1Experimental setup of activity monitors. A: IDEEA; B: Step-N-Tune; C: Activ4Life (attached directly to skin).
Group mean step counts for each measure.
| Measure | Control group | RA patients |
|---|---|---|
| Gold standard | 221.75 (19.21) | 159.50 (38.65) |
| Step-N-Tune | 213.75 (57.42) | 86.27 (77.57) |
| A4L | 247.17 (21.68) | 177.27 (58.04) |
| IDEEA | 184.67 (61.49) | 72.92 (69.65) |
Values indicate group average for each measure. Gold standard: mean of 3 counts from half speed video replay; A4L: Activ4Life; SD: standard deviation.
Fig. 2Regression based Bland–Altman plots illustrating agreement between different activity monitors and the gold standard. Solid line: mean agreement; dotted lines: 95% limits of agreement.
Regression based Bland–Altman 95% limits of agreement and estimated mean differences for step counts of each of three automated measures.
| Device | |||
|---|---|---|---|
| Step-N-Tune | A4L | IDEEA | |
| RA patients | |||
| Variation in bias, per step | − 1.11x | − 0.38x | − 1.53x |
| Width of 95% LOA | 169.74 | 52.90 | 225.00 |
| Bias (95% LOA): 100 steps | 110.59 (25.72 to 195.46) | 6.12 (20.33 to 32.57) | 127.84 (15.34 to 240.34) |
| 150 steps | 55.09 (− 29.78 to 139.96) | − 12.88 (− 39.33 to 13.57) | 51.34 (− 61.16 to 163.84) |
| 200 steps | − 0.41 (− 85.28 to 84.46) | − 31.88 (− 58.33 to − 5.43) | − 25.16 (137.66 to 87.34) |
| Controls | |||
| Variation in bias, per step | 0.011x | − 0.13x | − 0.05x |
| Width of 95% LOA | 22.64 | 44.16 | 61.52 |
| Bias (95% LOA): 200 steps | − 6.51 (− 17.83 to 4.81) | − 20.75 (− 42.83 to 1.33) | 21.47 (− 9.29 to 52.23) |
Bias = estimated mean difference between gold standard and test device; LoA = limits of agreement; x = mean of gold standard and test device.
Estimates are only provided for controls at 200 steps because there were no observed data at lower step counts.
Fig. 3Regression based Bland–Altman plots for temporal and spatial gait parameters. Solid line: mean agreement; dotted lines: 95% limits of agreement.